BACKGROUND OF THE INVENTION
Field of the Invention
[0001] The present invention relates to high frequency excited electrodeless discharge lamps,
and more particularly to improvements in the composition of the fill materials thereof
which, when excited, emit light that is particularly rich in the near ultraviolet
light range.
Description of the Prior Art
[0002] Near ultraviolet light sources which are often used for processes involving photochemical
reactions, such as photoengraving, have commonly comprised high pressure metal vapour
electric discharge lamps which have a pair of discharge electrodes idsposed within
the elevelope thereof. Such electric discharge lamps have generally comprised fill
materials including halides of gallium, etc., and thus have been called metal halide
lamps. This type of conventional metal halide lamps is disadvantageous, however, in
that the stabilization time, i.e., the length of time that the lamp requires to attain
the stable state of light emission after it is turned on, is relatively long, i.e.,
as long as about three minutes. Thus, when this type of conventional metal halide
lamps is used in the photoengraving process in which exposure and preparation steps
follow one after another at short intervals of one minuit, the lamps cannot be turned
off during the preparation steps between the exposure steps.
[0003] It has thus been necessary to keep the lamp continuously turned on behind a shutter
during all of the operations of the photoengraving process, opening the shutter during
the exposure steps in which the light is required. This causes much loss of electric
power. Thus, the so-called instant stabilization type ultraviolet light sources have
been much needed. Further, conventional metal halide lamps constitute high electrical
loads and the life thereof has been limited to about a thousand hours. This relatively
short life is due, for example, to the stains originating from the electrodes which
accumulate on the inner surface of the envelope of the lamp.
[0004] Thus, light sources have already been proposed in which electrodeless lamps are exciting
by high frequency waves, especially microwaves. These electrodeless discharge lamps
enjoy longer life than the conventional lamps with discharge electrodes, because a
main factor limiting the life of the conventional lamps has been the comsumption of
the electrodes and the stains resulting therefrom. A further advantage of the electrodeless
lamp is that there is no thermal loss at the discharge electrodes, and that it is
easier to apply greater electric power to the lamp from the time of turn-on, because
the impedance of the discharge in the electrodeless lamps vaires little from the time
it is turned on till it attains the stable state. Also, the stabilization time of
the electrodeless lamps is shorter because the electric discharge thereof concentrates
near the inner surface of the envelope of the lamp.
[0005] Although the electrodeless discharge lamps have many advantages as above described,
they have not been satisfactory as near ultraviolet light sources. That is, the conventional
fill materials thereof did not give enough light-output in the near ultraviolet range,-especially
in the range of 350 to 450 nm length. Such conventional fill materials are disclosed,
for example, in U.S. patent 4,001,632 issued to Haugsjaa et al. on Jan. 4, 1977 as
examples I through III in column 5 thereof.
[0006] The mechanism of light emission, however, is substantially the same in the electrodeless
discharge lamps as in the conventional discharge lamps having discharge electrodes.
Namely, the light emitting metal contained in the fill material sealed in the envelope
of the lamp is vapourized and excited by the high frequency waves to emit light. Thus,
when higher light emission is required in a particular wave length range, fill materials
comprising substantially the same kind of light emitting metals must be sealed in
the envelope in the electrodeless lamps as in the conventional lamps having discharge
electrodes. The light emission in the case of the conventional metal halide lamps
having discharge electrodes, however, concentrates near the axis between the discharge
electrodes, which are situated at the two end portions of the envelope, in contrast
to the case of the electrodeless discharge lamps in which the light emission extends
to the meighbourhood of the inner surface of the envelope even when the vapour pressure
within the envelope of the lamp is high.
[0007] Thus, the amounts of fill materials which are suitable for the conventional metal
halide lamps and which are suitable for the electrodeless lamps are different even
when fill materials including the same kind of metals are used.
SUMMARY OF THE INVENTION
[0008] Thus, the object of the invention is to provide an improved high frequency excited
electrodeless lamp which emits light that is particularly rich in the near ultraviolet
range, especially in the range of 350 to 450 nm wavelength. More particularly, the
present invention contemplates providing fill materials for the high frequency excited
electrodeless discharge lamps which give sufficient light emission in the near ultraviolet
range, especially in the 350 to 450 nm wave length range.
[0009] Thus, according to one.aspect of the present invention, the fill material sealed
within the light transmitting envelope of the electrodeless lamp comprises a rare
gas, mercury, a halogen, and a metal selected from the group consisting of iron, nickel,
cobalt, palladium, and the mixtures thereof. The fill material comprises, per one
cubic centimeter of the volumetric content of the envelope of the lamp, mercury in
an amount from 7 to 55 preferably from 17.6 to 41.3 and more preferably around 25,
micromoles, the metal selected from the group in the total amount from 0.1 to 2.3,
preferably from 0.38 to 1.91 and more preferably from 0.5 to 1, micromoles, and the
halogen in a total amount ranging from 0.2 to 6.2 micromoles in terms of atoms or
irons thereof. Preferably, the amount of halogen atoms measured in terms of micromoles
per one cubic centimeter of the content of the envelope exceeds twice the amount of
the metal selected from the group measured in the same terms, by an amount ranging
from 0.02 to 2.0 micromoles. That is, it is preferable that there is enough halogen
for changing all the metal selected from the group into the halide thereof. It is
further preferred that the raregas is present in the envelope at a pressure ranging
from 10 to 200, preferably from 20 to 150 and more preferably from 30 to 130, torr.
[0010] The micromole unit used in the above measurements is equal to 10
-6 moles. The mole unit is the SI unit which is equivalent to the former corresponding
units such as gram-atom or gram-molecule.
[0011] According to another aspect of the present invention, the fill material sealed within
the envelope of the lamp comprises a rare gas, mercury, a halogen, and a rare earth
metal selected from the group consisting of dysprosium, holmium, thulium, scandium,
and the mixtures thereof. It is preferred that the fill material comprise, for each
one cubic centimeter of the content of the envelope, mercury in an amount ranging
from 5 to 55 preferably from 17.6 to 53, micromoles, a rare earth metal in a total
amount ranging from 0.05 to 0.6 preferably from 0.13 to 0.39 and more preferably around
0.25, micromoles, and halogen in a total amount ranging from 0.15 to 0.62 micromoles
in terms of atoms thereof. It is preferred that the amount of halogen atoms atoms
measured in terms of micromoles per one cubic centimeter of the volumetric content
of the envelope exceed three times the amount of rare earth metal measured in the
same terms.
[0012] That is, it is preferred that there be enough halogen to - combine with all the rare
earth metal present in the envelope to form the halide thereof.
[0013] In both aspects of the invention, the halogen may be iodine, bromine, or a mixture
thereof, and the rare gas may be argon.
BRIEF DESCRIPTION OF THE DRAWINGS
[0014] Further details of the present invention will become more apparent from the following
detailed description of the preferred embodiments, taken in conjunction with the accompanying
drawings, in which:
Fig. 1 is a schematic cross-sectional view of the microwave generating device in which
the electrodeless discharge lamp according to the present invention may be disposed;
Fig. 2 is a cross-sectional view of an electrodeless lamp accoring to the present
invention;
Fig. 3 shows the curve representing the relationship between the near ultraviolet
output of an electrodeless lamp and the iron iodide content sealed in the envelope
thereof, wherein the fill material sealed in the envelope of the lamp comprises the
elements iron, iodine, mercury, and argon, and wherein the mercury and argon contents
are fixed;
Fig. 4 shows the curve representing the relationship between the near ultraviolet
output of an electrodelsss lamp and the mercury content sealed in the envelope thereof,
wherein the fill material sealed in the envelope comprises the elements iron, iodine,
mercury, and argon, and wherein the iron, iodine, and argon contents are fixed;
Fig. 5 shows the curve representing the relationship between the near ultraviolet
output of an electrodeless lamp and the dysprosium iodide content sealed in the envelope
thereof, wherein the fill material sealed in the envelope comprises the elements dysprosium,
iodine, mercury, and argon, and wherein the mercury and argon contents are substantially
fixed;
Fig. 6 shows the curve representing the relationship between the near ultraviolet
light output of an electrodeless lamp and the mercury content sealed in the envelope
thereof, wherein the fill material thereof comprises the elements dysprosium, iodine,
mercury, and argon, and the dysprosium, iodine, and argon contents are fixed.
Fig. 7 shows the curves representing the variations of the near ultraviolet light
output and the starting time against the mole fraction of bromine with respect to
the total molar content of bromine and iodine included in the fill material of an
electrodeless lamp.
[0015] In the drawings like reference numerals represent like components.
DETAILED DESCRIPTION OF THE PREFERRED EMBODIMENTS
[0016] Referring now to Fig. 1 and 2 of the drawings, a construction of an electrodeless
discharge lamp according to the present invention is described, together with that
of a microwave generating device in which the lamp according to the present invention
may be disposed.
[0017] The microwave generating device of Fig. 1 comprises a magnetron 1 generating a microwave
of 2450 MHz and having an output power of 700 W. The magnetron 1 is disposed at an
end portion of a wave guide 3, and has a magnetron antenna 2 from which the microwave
is radiated into the wave guide 3. The wave guide 3 opens into a cavity 4 enclosed
by cavity wall 5 and a metallic mesh plate 11 at a microwave feeder opening 6. The
cavity wall 5 is formed of a substantially semispherical aluminum plate and the inner
surface thereof forms a light reflecting surface. The metallic mesh plate 11 is a
stainless steel mesh plate manufactured by the etching method, and transmits about
85 percent of light therethrough but not the microwave generated by the magnetron
11. And electrodeless lamp 7 comprieses a spherical envelope 7a formed of light transmitting
quartz and a pair of rod- shaped projections 7b and 7c formed of the same material
and integral therewith. The spherical envelope 7a has a thickness of 0.5 mm and an
inner diameter of 30 mm, thus enclosing a spherical space of about 14.1 cm3"in which
an inert gas, metals, etc., are sealed in a certain composition according to the present
invention, as will be described in detail hereinafter. The projections 7b and 7c have
a length of 10 mm and a diameter of 3 mm, and the electrodeless lamp 7 is supported
at these projections 7b and 7c by supporting members (not shown) formed on the cavity
wall 5. A ventilator fan 8 disposed at an end of a ventilating duct 9 introduces cooling
air into the duct 9 from outside the housing 12 which accomodates the magnetron 1,
the wave guide 3, etc., and thus cools the magnetron 1 and the electrodeless lamp
7.
[0018] The operation of the microwave generating device of Fig. 1 will be described, together
with that of an electrodeless lamp according to the present invention. The microwave
generated by the magnetron 1 is radiated into the wave guide 3 from the magnetron
antenna 2, and then is propagated through the wave guide 3 and radiated into the cavity
4 from the feeder opening 6, thereby forming a microwave electromagnetic field in
the cavity 4. Thus, the electrodeless lamp 7 is placed in the microwave electromagnetic
field established in the cabity 4, and the inert gas, which is sealed in the lamp
7, for starting the electric discharge within the lamp 7, begins to discharge, thereby
beating the inner surface of the envelope 7a of the electrodeless lamp 7. Thus, the
metals deposited on the inner surface of the envelope 7a of the lamp 7 begin to evaporate,
filling the space within the envelope 7a with metal vapor. Thus, the electric discharge
within the envelope 7a is now carried out, for the main part thereof, by the metal
vapor, and is stabilized when the metal vapor discharge takes place, the metal vapor
emits light having the emission spectra characteristic of the metals. This light emitted
from the metal vapors is utilized as a light source. Further, in order to effectively
utilize the light emitted from the electrodeless lamp 7, the inner surface of the
cavity wall 5 is made light reflecting, and the front of the cavity 4 is covered by
the metallic mesh plate 11 which transmits light but not microwaves. Thus substantially
all the light emitted from the electrodes lamp 7 is radiated forward through the mesh
plate 11. Further, as it is necessary to remove heat generated in the magnetron 1
and the lamp 7, the ventilating fan 8 takes in the outside air, which then blows through
the duct 9, the opening 10, the wave guide 3, the feeder opening 6, and the cavity
7, and is exhausted from the cavity 7 through the mesh plate 11.
[0019] We have conducted a series of experiments to determine the optimum composition of
the fill material of an electrodeless discharge lamp which can be used as near an
ultraviolet light source. In all the experiments described hereinbelow, the electrodeless
discharge lamp 7 of Fig. 2 having the physical construction and dimensions described
hereinabove was used and placed in the microwave generating device of Fig. 1. Thus,
the volumetric content of the envelope 7a is 14.1 cubic centimeters, and it should
be understood that when the amounts of substances contained in the fill material of
the lamp 7 are expressed in terms of micromoles per cubic centimeter, the actual molar
amounts of the substances contained in the fill material are obtained by multiplying
the values expressed in terms of micromoles per cubic centimeter by the factor of
14.1 cubic centimeters.
[0020] Further, it should be obvious that when electrodeless discharge lamps having a larger
or smaller volumetric content are used, the actual amounts of the substances of the
fill material should be increased or decreased in proportion to the volumetric content
of the envelope. The physical construction and the dimensions of the electrodeless
lamp 7 of Fig. 2 were described hereinabove only for exemplary purposes, and the scope
of the invention is not limited to the particular form or dimensions of the lamps
of Fig. 2.
[0021] In all the experiments described hereinbelow,
f except for the last one of Fig. 7, the light output of the lamp 7 in the near ultraviolet
range of 355 to 425 nm has been measured on an arbitary scale, in which the electrodeless
discharge lamp formerly developed in our laboratory and including gallium as the light
emitting metal scored a value 65.
[0022] Referring now to Figs. 3 and 4 of the drawings, the first series of experiments in
which the fill material of the lamp 7 comprising the elements iron, iodine, mercury,
and argon, is described.
[0023] Fig. 3 shows the dependence of the light output on the iron iodide content of the
fill material. Namely, fixed amounts of argon and mercury, i.e., argon at pressure
of 100 torr as the starter rare gas and mercury in an amount of 120 mg (or 42 micromoles
per cubic centimeter of the content of the envelope 7a) as the buffer gas were sealed
in the envelope 7a of the lamp 7. Further, a variable amount of iron iodide (FeI
2) was sealed in and the dependance of the light output of the lamp 7 on the amount
of iron iodide was measured. Thus, the fill material used in this experiment comprised
fixed amounts of the elements mercury and argon, and variable amounts of the elements
iron and iodine. The amount of iodine in terms of moles was two times that of iron.
The elements iron and iodine could of course be sealed in the form of metallic iron
and mercury iodide (HgI2), keeping the ratio of the amount of these elements in terms
of moles at 1
:2.
[0024] As shown in Fig. 3, the light output in the near ultraviolet range increases rapidly
at first with increase in the amount of iron iodide sealed in the envelope 7a, and
reaches a maximum when the amount of iron iodide is between 0.5 and 1.0 micromoles
per cubic centimeter. An amount of iron iodide between 0.1 and 2.3 micromoles per
cubic centimeter is practically feasible. When the amount of iron iodide sealed in
the envelope 7a is less than 0.1 micromoles, it is difficult to seal in the precise
predetermined amount of iron iodide due to measurement errors and fluctuations in
the parameters in the manufacture of the lamp, and thus the near ultraviolet light
outputs of the product lamps vary considerably from each other. Further when the amount
of iron iodide sealed in the envelope 7a exceeds 2.3 micromoles per cubic centimeter,
the discharge within the envelope 7a becomes unstable and fluctuates, presenting a
striped pattern therein.
[0025] Fig. 4 shows the dependance of the near ultraviolet light output of the electrodeless
lamp 7 upon the mercury content of the fill material thereof. The argon gas was s
paled at fixed pressure of 100 torr as the starter gas, and iron in the fixed amount
of 0.63 micromoles per cubic centimeter was sealed in the envelope 7a as the light
emitting metal, together with mercury iodide in the fixed amount of 0.62 micromoles
per cubic centimeter. The amounts of these substances were fixed at these values while
the amount of mercury sealed in the envelope was changed and the dependance of the
light output of the lamp 7 on the amount of mercury was measured. Thus, in this experiment
of Fig. 4, the fill material comprised the elements iron and iodine in the fixed amounts
of 0.63 and 1.24 micromoles per cubic centimeter, respectively, and also a fixed amount
of argon at a pressure of 100 torr in the envelope 7a.
[0026] As shown in Fig. 4, the light output of the lamp 7 increases at first rapidly with
an increase in the amount of mercury sealed in the envelope 7a, and reaches a maximum
when the amount of mercury is at about 25 micromoles per cubic centimeter. The light
output decreases gradually when the amount of mercury is increased beyond about this
value. The amount of mercury in the range of from 7 to 55 micromoles is practically
feasible. The reason is that when the amount of mercury is less than 7 micromoles
per cubic centimeter, the light output is not sufficient, and, on the other hand,
when it exceeds 55 micromoles per cubic centimeter, the light emission in the envelope
7a presents a striped pattern and becomes unstable with fluctuation.
[0027] Further experiments were conducted changing the pressure of argon gas in the envelope
7a of the lamp 7, the fill materials of which comprised the same elements as in the
case of the experiments of Figs. 3 and 4. Namely, iron in the fixed amount of 0.63
micromoles per cubic centimeter and mercury iodide in the fixed amount of 0.62 micromoles
per cubic centimeter were sealed in the envelope 7a, and the pressure of argon gas
in the envelope 7a was changed from 1, through 5, 10, 40, 100, and 200, to 300 torr.
[0028] When the pressure of argon gas is at 1 torr, the discharge within the envelope extinguished
before it reached the stable state, and when the pressure of argon gas is at 300 torr,
the lamp 7 did not start light emission. Thus, it was found preferable.to limit the
pressure of argon in the envelope 7a within the range of from 10 to 200 torr. As the
result of further experiments, it was found that a more preferalbe range of the pressure
of argon in the envelope 7a in the case where the fill material comprises iron, iodine,
mercury and argon is between 20 and 150 torr, and a still more preferable range thereof
is between 30 and 130 torr.
[0029] With regard to the iodine content in the fill material, an amount thereof necessary
to form the sufficient amount of iron iodide should be sealed in the envelope 7a.
When the iodine is sealed in the envelope 7a in the form of mercury iodide, the maximum
amount of iodine which can be sealed in is 1.25 mg, or 6.2 micromoles per cubic centimeter.
The reason of this is that when the amount of iodine in the fill material exceeds
this maximum of 6.2 micromoles per cubic centimeter the light emission in the envelope
7a becomes uneven, and the discharge within the envelope 7a unstable with fluctuations.
Thus, the amount of iodine in the fill material should not be less than 0.2 micromoles
per cubic centimeter, which is necessary to form the minimum permissible amount of
iron iodide of 0.1 micromoles per cubic centimeter, and not more than 6.2 micromoles
per cubic centimeter. Further, it is preferred that the amount of iodine measured
in terms of micromoles per cubic centimeter exceeds 2 times the amount of iron measured
in the same terms. Namely, it is preferable that the amount of iodine exceeds the
amount thereof which is necessary to combine with all the iron present in the fill
material. The preferable excess amount of iodine is from 0.02 to 0.2 micromoles per
cubic centimeter.
[0030] Thus, in view of the experiments described above and also in view of further experiments,
we conclude as follows with regard to the fill materials of an electrodeless discharge
lamp which comprises a rare gas, mercury, a halogen, and a metal selected from the
group consisting of iron, nickel, cobalt, palladium, and the mixture thereof.
[0031] The fill material should comprise, per one cubic centimeter of the content of the
envelope of the lamp mercury in an amount from 7 to 55 preferably from 17.6 to 41.3
and more preferably around 25, micromoles, the metal selected from the group in overall
amount from 0.1 to 2.3, preferably from 0.38 to 1.91 and more preferably from 0.5
to 1, micromoles, and the halogen in an overall amount ranging from 0.2 to 6.2 micromoles.
Preferably, the amount of halogen measured in terms of micromoles per one cubic centimeter
of the content of the envelope exceeds twice the amount of the metal selected from
the group measured in the same term by an amount ranging from 0.02 to 2.0 micromoles.
That is, it is preferable that there is enough halogen for changing all the metal
selected from the group into the halide thereof. It is further preferred that the
rare gas is present in the envelope at a pressure ranging from 10 to 200 preferably
from 20 to 150 and more preferably from 30 to 130, torr.
[0032] The micromole unit used in the above measurements of the substances involved is equal
to 10 moles. The mole unit is the SI unit which is equivalent to the former corresponding
units such as gram-atom or gram-molecule units.
[0033] Referring now to Figs. 5 and 6 of the drawings, the second series of experiments
in which the fill material of the lamp 7 comprised the elements dysprosium, iodine,
mercury, and argon, is described.
[0034] Fig. 5 shows the dependence of the light output on the dysprosium iodide content
of the fill material. Namely, fixed amounts of argon and mercury, i.e., argon at a
pressure of 100 torr as the starter rare gas and mercury in the amount of 100 mg as
the buffer gas were sealed in the envelope 7a of the lamp 7. Further, variable amounts
of dysprosium and mercury iodide were sealed in and the dependence of the light output
of the lamp 7 on the amount of dysprosium iodide (DyI
3) was measured. Namely, variable amounts of dysprosium and mercury iodide were sealed
in the envelope 7a, while keeping the ratio thereof in terms of moles at 1:1.5. Thus,
the ratio of the amounts of dysprosium and iodine in terms of moles was kept at 1:3,
i.e., there was just enough iodine to conbine with all the dysprosium to form dysprosium
iodide. Thus, the fill material used in this experiment comprised substantially fixed
amounts of the elements mercury and argon, and variable amounts of the elements iron
and iodine.
[0035] As shown in Fig. 5, the light output in the near ultraviolet range increases rapidly
at first with an increase in the amount of dysprosium iodide sealed in the envelope
7a, and reaches a maximum when the amount of dysprosium iodide is about 0.25.micromoles
per cubic centimeter. An amount of dysprosium iodide between 0.05 and 0.6 micromoles
per cubic centimeter is practical. When the amount of dysprosium iodide sealed in
the envelope 7a is outside this range, the light output does not improve comspicuously.
[0036] Fig. 6 shows the dependence of the near ultraviolet light output of the electrodeless
lamp 7 upon the mercury content of the fill material thereof. The argon gas was sealed
at fixed pressure of 100 torr as the starter gas, and dysprosium in the fixed amount
of 0.26 micromoles per cubic centimeter was sealed in the envelope 7a as the light
emitting metal, together with mercury iodide in the fixed amount of 0.39 micromoles
per cubic centimeter. The amounts of these substances were fixed at these values,
while the amount of mercury sealed in the envelope was changed and the dependence
of the light output of the lamp 7 on the amount of mercury was measured. Thus, in
this experiment of Fig. 6, the fill material comprised the elements dysprosium and
iodine in the fixed amounts of 0.26 and 0.78 micromoles per cubic centimeter, respectively,
and also a fixed amount of argon at a pressare of 100 torr in the envelope 7a.
[0037] As shown in Fig. 6, the light output of the lamp 7 increases rapidly at first with
an increase in the amount of mercury sealed in the envelope 7a, and saturated when
the amount of mercury is at about 50 micromoles per cubic centimeter. The light output
is substantially constant when the amount of mercury is increased beyond about this
value. The amount of mercury in the range of from 5 to 55 micromoles is practically
feasible. The reason is that when the amount of mercury is less than 5 micromoles
per cubic centimeter, the light output is not improved sufficiently, and, on the other
hand, when it exceeds 55 micromoles per cubic centimeter, the light emission in the
envelope 7a presents a striped pattern and becomes unstable with fluctuation.
[0038] With regard to the iodine content in the fill material, an amount thereof necessary
to form sufficient amount of dysprosium iodide DyI
3 should be sealed in the envelope 7a. When the iodine is sealed in the envelope 7a
in the form of mercury iodide, the maximum amount of iodine which can be sealed in
is 1.25 mg, or 6.2 micromoles per cubic centimeter. The reason of this is that when
the amount of iodine in the fill material exceeds this maximum of 6.2 micromoles per
cubic centimeter, the light emission in the envelope 7a becomes uneven and the discharge
within the envelope 7a becomes unstable with fluctuations. Thus, the amount of iodine
in the fill material should not be less than 0.15 micromoles per cubic centimeter,
which is necessary to form the minimum permissible amount of dysprosium iodide of
0.05 micromoles per cubic centimeter, and not more than 6.2 micromoles per cubic centimeter.
Further, it is preferred that the amount of iodine measured in terms of moles per
cubic centimeter exceeds 3 times the amount of dysprosium measured in the same terms.
Namely, it is preferable that the amount of iodine exceeds the amount thereof which
is necessary to combine with all the dysprosium present in the fill material.
[0039] Thus, in view of the experiments described above and also in view of further experiments,
we conclude as follows with regard to the fill materials of an electrodeless discharge
lamp which comprise a rare gas, mercury, a halogen, and a metal selected from the
group consisting of,dysprosium, holmium, thulium, scandium, and the mixtures thereof.
[0040] The fill material should comprise for each one cubic centimeter of the content of
the envelope, mercury in an amount ranging from 5 to 55 preferably exceeding 17.5,
micromoles, a rare earth metal in an overall amount ranging from 0.05 to 0.6 preferably
from 0.13 to 0.39 and more preferably around 0.25, micromoles, and halogen in an overall
amount ranging from 0.15 to 0.62 micromoles. It is preferred that the amount of halogen
measured in terms of micromoles per one cubic centimeter of the volumetric content
of the envelope exceed three times the amount of the rare earth metal measured in
the same terms. That is, it is preferred that there be enough halogen to combine with
all the rare earth metal present in the envelope to form the halide thereof.
[0041] Referring to Fig. 7 of the drawings, a third series of experiments is now described.
[0042] In the envelope 7a of the electrodeless discharge lamp 7, a fill material was sealed
which comprises mercury in the amount of 100 mg, iron in the amount of 0.3 mg, and
argon as the rare gas at the pressure of 60 torr. The fill material further comprised
3 mg of mercury iodide in the first experimental example. In the second experimental
example, the fill material comprised 2 mg of mercury iodide and 1 mg of mercury bromide.
In the third experimental example, the fill material comprised 1 mg of mercury iodide
and 2 mg of mercury bromide. In the fourth experimental example, the fill material
comprised 3 mg of mercury bromide. Thus, in addition to the fixed amounts of mercury,
iron, and argon, all four examples comprised, mercury iodide and/or mercury bromide
in the fixed total amount of 3 mg, but in variable mole fractions thereof. The lamp
7 was placed in the microwave generating device of Fig. 1, and the light output thereof
in the near ultraviolet wave length range of 350 to 450 nm and the starting time,
i.e., the time required by the lamp after it is turned on to attain 80 per cent of
the light output of stable light emission state thereof were measured.
[0043] The abscissa in Fig. 7 represents the mole fraction of bromine with respect to the
total molar content of iodine and bromine in percent, i.e.,

wherein M1and M
2 are the amounts of bromine and iodine respectively, in terms of moles. The ordinate
of Fig. 7 represents the light output of the lamp in an arbitary scale and the starting
time in seconds. The solid line T represents the starting time and the dotted line
P represents the light output.
[0044] As shown in Fig. 7, the starting times of the lamp 7 in the cases where the fill
material comprises solely iodine or bromine, respectively, are 13.5 and 16.2 seconds.
The starting time is reduced when both bromine and iodine are included in the fill
material, and when the mole fraction of bromine is chosen between 10 and 77.5 per
cent, the starting time is reduced under 10 seconds. The reason why the starting time
can be reduced by utilising two halogen elements instead of only one is that the vapour
pressure of the halides of the light emitting metal comes to be the sum of the vapour
pressures of two kinds of halides of the metal, and thus the appropriate vapour pressure
of the halides of the metal is reached at a time when the temperature of the surface
of the envelope 7a of the lamp 7 is lower than that at the stable state thereof.
[0045] As shown further in Fig. 7, the light output in the wave length range of 350 to 450
nm is increased when both iodine and bromine are included in the fill material, as
compared to the case in which only iodine or bromine is included. The light output
of the lamp the fill material of which includes only mercury as the light emitting
metal was 31.0 in this arbitary scale, and thus the maximum light output of the lamp
used in this experiment scored 3.2 times as much as that conventional lamp.
[0046] In the experiments described above, the light emitting metal was iron, and the halogen
iodine and bromine. By further experiments, however, it was confirmed that the starting
time is reduced when combination of halogens other than bromine and iodine are used,
i.e., when, for example, the combination of iodine and chlorine, of bromine and chlorine,
or of iodine, bromine, and chlorine is used. Further, the starting time of the lamp
7 is expected to be reduced when light emitting metal other than iron is used.
Example I
[0047] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprises iron in the amount
of 0.5 mg or 0.63 micromoles per cubic centimeter, mercury iodide (HgI
2) in the amount of 4 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount
of 118 mg, and argon at the pressure of 100 torr. Thus, the fill material comprised
mercury in the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The
amount of iodine in terms of atoms was 1.24 micromoles per cubic centimeter which
was less than two times the amount of iron measured in the same terms, by an amount
of 0.02 micromoles per cubic centimeter. Namely, there was a shortage of 0.02 micromoles
per cubic centimeter of iodine to combine with all the iron present in the fill material.
[0048] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 206 in an arbitary scale,
which is 4.12 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 17.0 seconds.
Example II
[0049] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised palladium in the amount
of 1.0 mg, iodine in the amount of 4 mg, mercury in the amount of 118 mg, and argon
at the pressure of 100 torr.
[0050] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 103 in an arbitary scale,
which is 2.1 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 19.0 seconds.
Example III
[0051] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.3 mg or 0.38 micromoles per cubic centimeter, mercury iodide in the amount of
2.5 mg, or 0.39 micromoles per cubic centimeter, mercury in the amount of 99 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 0.78 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.01 micromoles per cubic centimeter.
Namely, there was an excess of 0.01 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0052] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 196 in an arbitary scale,
which is 3.92 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 16.5 seconds.
Example IV
[0053] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimer, mercury in the amount of 98 mg, and
argon at the pressure of 100 mg, or 35.3 micromoles per cubic centimer. The amount
of iodine atoms was 1.72 micromoles per cubic centimer which exceeded two times the
amount of iron measured in the same terms, by an amount of 0.06 micromoles per cubic
centimeter. Namely, there was an excess of 0.06 micromoles per cubic centimeter of
iodine to combine with all the iron present in the fill material.
[0054] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 216 in an arbitary scale,
which is 4.32 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 17.5 seconds.
Example V
[0055] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 1.0 mg or 1.29 micromoles per cubic centimeter, mercury iodide in the amount of
8.0 mg, or 1.24 micromoles per cubic centimeter, mercury in the amount of 97 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 2.48 micromoles per cubic centimeter which was less than two times the amount
of iron measured in the same terms, by an amount of 0.08 micromoles per cubic centimeter.
Namely, there was a shortage of 0.08 micromoles per cubic centimeter of iodine to
combine with all the iron present in the fill material.
[0056] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 210 in an arbitary scale,
which is 4.20 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0057] The starting time on the other hand was 18.0 seconds.
Example VI
[0058] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 1.5 mg or 1.91 micromoles per cubic centimeter, mercury iodide in the amount of
12.0 mg, or 1.87 micromoles per cubic centimeter, mercury in the amount of 95 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of
iodine atoms was 3.74 micromoles per cubic centimeter which was less than two times
the amount of iron measured in the same terms, by an amount of 0.08 micromoles per
cubic centimeter. Namely, there was a shortage of 0.08 micromoles per cubic centimeter
of iodine to combine with all the iron present in the fill material.
[0059] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 196 in an arbitary scale,
which is 3.92 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 18.0 seconds.
Example VII
[0060] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.24 micromoles per cubic centimeter which was less than two times the amount
of iron measured in the same terms, by an amount of 0.42 micromoles per cubic centimeter.
Namely, there was a shortage of 0.42 micromoles per cubic centimeter of iodine to
combine with all the iron present in the fill material.
[0061] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 206 in an arbitary scale,
which is 4.12 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 17.0 seconds.
Example VIII
[0062] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
7.0 mg, or 1.10 micromoles per cubic centimeter, mercury in the amount of 97 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, 35.3 or micromoles per cubic centimeter. The amount of iodine
atoms was 2.20 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.54 micromoles per cubic centimeter.
Namely, there was an excess of 0.54 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0063] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 210 in an arbitary scale,
which is 4.2 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 16.5 seconds.
Example IX
[0064] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
10.0 mg, or 1.56 micromoles per cubic centimeter, mercury in the amount of 96 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of
iodine atoms was 3.12 micromoles per cubic centimeter which exceeded two times the
amount of iron measured in the same terms, by an amount of 0.73 micromoles per cubic
centimeter. Namely, there was an excess of 0.73 micromoles per cubic centimeter of
iodine to combine with all the iron present in the fill material.
[0065] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 206 in an arbitary scale,
which is 4.12 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 15.0 seconds.
Example X
[0066] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
15.0 mg, or 2.34 micromoles per cubic centimeter, mercury in the amount of 93 mg,
and argon at the pressure of 100torr. Thus, the fill material comprised mercury in
the total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of
iodine atoms was 4.68 micromoles per cubic centimeter which exceeded two times the
amount of iron measured in the same terms, by an amount of 3.01 micromoles per cubic
centimeter. Namely, there was an excess of 3.01 micromoles per cubic centimeter of
iodine to combine with all the iron present in the fill material.
[0067] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 190 in an arbitary scale,
which is 3.80 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0068] The starting time on the other hand was 15.0 seconds.
Example XI
[0069] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2,the fill material was sealed which comprised iron in the amount of
0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of 5.5
mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 48 mg, and argon
at the pressure of 100 torr. Thus, the fill material comprised mercury in the total
amount of 50 mg, or 17.6 micromoles per cubic centimeter. The amount of iodine atoms
was 1.72 micromoles per cubic centimeter which exceeded two times the amount of iron
measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess 0.06 of micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0070] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 203 in an arbitary scale,
which is 4.06 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 13.0 seconds.
Example XII
[0071] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which compried iron in the amount of
0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of 5.5
mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 68 mg, and argon
at the pressure of 100 torr. Thus, the fill material comprised mercury in the total
amount of 70 mg, or 24.7 micromoles per cubic centimeter. The amount of iodine atoms
was 1.72 micromoles per cubic centimeter which exceeded two times the amount of iron
measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0072] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 212 in an arbitary scale,
which is 4.24 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 16.0 seconds.
Example XIII
[0073] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 88 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 90 mg, or 31.8 micromoles per cubic centimeter. The amount of iodine
atoms was 1.72 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0074] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 215 in an arbitary scale,
which is 4.30 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0075] The starting time on the other hand was 17.5 seconds.
Example XIV
[0076] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 20 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.72 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0077] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 220 in an arbitary scale,
which is 4.4 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 22.0 seconds.
Example XV
[0078] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 40 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.72 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0079] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 218 in an arbitary scale,
which is 4.36 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 19.5 seconds.
Example XVI
[0080] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 80 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 per-cubic centimeter. The amount of iodine atoms was
1.72 micromoles per cubic centimeter which exceeded two times the amount of iron measured
in the same terms, by an amount of 0.06 micromoles per cubic centimeter. Namely, there
was an excess of 0.06 micromoles per cubic centimeter of iodine to combine with all
the iron present in the fill material.
[0081] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 216 in an arbitary scale,
which is 4.32 as much as that of the lamp 7 the fill material of which includes only
mercury. The starting time on the other hand was 18.0 seconds.
Example XVII
[0082] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 120 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.72 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0083] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 214 in an arbitary scale,
which is 4.28 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 16.0 seconds.
Example XVIII
[0084] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sea;ed which comprised iron in the amount
of 0.65 mg or 0.83 micromoles per cubic centimeter, mercury iodide in the amount of
5.5 mg, or 0.86 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 150 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.72 micromoles per cubic centimeter which exceeded two times the amount
of iron measured in the same terms, by an amount of 0.06 micromoles per cubic centimeter.
Namely, there was an excess of 0.06 micromoles per cubic centimeter of iodine to combine
with all the iron present in the fill material.
[0085] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 213 in an arbitary scale,
which is 4.26 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 16.0 seconds.
Example XIX
[0086] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp-7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 118 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine in terms of atoms was 1.24 micromoles per cubic centimeter which exceeded three
times the amount of dysprosium measured in the same terms, by an amount of 0.46 micromoles
per cubic centimeter. Namely, there was an excess of 0.46 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0087] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. l, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 186 in an arbitary scale,
which is 3.72 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 22.0 seconds.
Example XX
[0088] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised scadium in the amount
of 0.3 mg, mercury iodide in the amount of 4 mg, mercury in the amount of 118 mg,
and argon at the pressure of 100 torr.
[0089] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 152 in an arbitary scale,
which is 3.04 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 20.0 seconds.
Example XXI
[0090] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.3 mg or 0.13 micromoles per cubic centimeter, mercury iodide in the amount of
2.0 mg, or 0.31 micromoles per cubic centimeter, mercury in the amount of 119 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine atoms was 0.62 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium measured in the same terms, by an amount of 0.23 micromoles per
cubic centimeter. Namely, there was an excess of 0.23 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0091] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 162 in an arbitary scale,
which is 3.24 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0092] The starting time on the other hand was 22.0 seconds.
Example XXII
[0093] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.9 mg or 0.39 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of l18 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine atoms was 1.24 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium mesasured in the same terms, by an amount of 0.07 micromoles
per cubic centimeter. Namely, there was an excess of 0.07 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0094] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 144 in an arbitary scale,
which is 2.88 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 17.0 seconds.
Example XXIII
[0095] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 1.2 mg or 0.52 micromoles per cubic centimeter, mercury iodide in the amount of
6.5 mg, or 1.015 micromoles per cubic centimeter, mercury in the amount of l17 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine atoms was 2.03 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium measured in the same terms, by an amount of 0.47 micromoles per
cubic centimeter. Namely, there was an excess of 0.47 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0096] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 96 in an arbitary scale,
which is 1.92 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0097] The starting time on the other hand was 15.5 seconds.
Example XXIV
[0098] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 48 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 50 mg, or 17.6 micromoles per cubic centimeter. The amount of iodine
atoms was 1.24 micromoles per cubic centimeter which exceeded three times the amount
of dysprosium measured in the same terms, by an amount of 0.46 micromoles per cubic
centimeter. Namely, there was an excess of 0.46 micromoles per cubic-centimeter of
iodine to combine with all the dysprosium present in the fill material.
[0099] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of . the stable state light emission, were measured.
The light output thus measured during the stable state scored 162 in an arbitary scale,
which is 3.24 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 15.0 seconds.
Example XXV
[0100] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 73 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 75 mg, or 26.5 micromoles per cubic centimeter. The amount of iodine
atoms was 1.24 micromoles per cubic centimeter which exceeded three times the amount
of dysprosium measured in the same terms, by an amount of 0.46 micromoles per cubic
centimeter. Namely, there was an excess of 0.46 micromoles per cubic centimeter of
iodine to combine with all the dysprosium present in the fill material.
[0101] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 170 in an arbitary scale,
which is 3.4 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0102] The starting time on the other hand was 18.0 seconds.
Example XXVI
[0103] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 98 mg, and
argon at the pressure of 100 torr. Thus, the fill material comprised mercury in the
total amount of 100 mg, or 35.3 micromoles per cubic centimeter. The amount of iodine
atoms was 1.24 micromoles per cubic centimeter which exceeded three times the amount
of dysprosium measured in the same terms, by an amount of 0.46 micromoles per cubic
centimeter. Namely, there was an excess of 0.46 micromoles per cubic centimeter of
iodine to combine with all the dysprosium present in the fill material.
[0104] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 182 in an arbitary scale,
which is 3.64 times as much as that of the lamp 7 the fill material of which includes
only mercury. The starting time on the other hand was 22.0 seconds.
Example XXVII
[0105] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
4.0 mg, or 0.62 micromoles per cubic centimeter, mercury in the amount of 148 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 150 mg, or 53.0 micromoles per cubic centimeter. The amount of
iodine atoms was 1.24 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium measured in the same terms, by an amount of 0.46 micromoles per
cubic centimeter. Namely, there was an excess of 0.46 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0106] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 194 in an arbitary scale,
which is 3.88 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0107] The starting time on the other hand was 23.0 seconds.
Example XXVIII
[0108] In the envelope 7a having the volumetric content of
14.1 cubic centimeters of the lamp 7 of Fig. 2, the fill material was sealed which
comprised dysprosium in the amount of
0.
6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of 2.0 mg,
or 0.31 micromoles per cubic centimeter, mercury in the amount of 119 mg, and argon
at the pressure of 100 torr. Thus, the fill material comprised mercury in the total
amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of iodine atoms
was 0.62
mi
cromoles per cubic centimeter which was less than three times the amount of dysprosium
measured in the same terms, by an amount of 0.16 micromoles per cubic centimeter.
Namely, there was a shortage of 0.16 micromoles per cubic centimeter of iodine to
combine with all the dysprosium present in the fill material.
[0109] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig.
1, and the light output in the wave length range of 350 to
425 nm and the starting time, i.e., the length of time that the lamp 7 required to attain
80 percent light emission of the stable state light emission, were measured. The light
output thus measured during the stable state scored 183 in an arbitary scale, which
is 3.66 times as much as that of the lamp 7 the fill material of which includes only
mercury. The starting time on the other hand was 22.0 seconds.
Example XXIX
[0110] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
6.0 mg, or 0.935 micromoles per cubic centimeter, mercury in the amount of 117 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine atoms was 1.87 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium measured in the same terms, by an amount of 1.35 micromoles per
cubic centimeter. Namely, there was an excess of 1.35 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0111] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 186 in an arbitary scale,
which is 3.72 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0112] The starting time on the other hand was 22.0 seconds.
Example xxx
[0113] In the envelope 7a having the volumetric content of 14.1 cubic centimeters of the
lamp 7 of Fig. 2, the fill material was sealed which comprised dysprosium in the amount
of 0.6 mg or 0.26 micromoles per cubic centimeter, mercury iodide in the amount of
12.0 mg, or 1.875 micromoles per cubic centimeter, mercury in the amount of 115 mg,
and argon at the pressure of 100 torr. Thus, the fill material comprised mercury in
the total amount of 120 mg, or 42.4 micromoles per cubic centimeter. The amount of
iodine atoms was 3.75 micromoles per cubic centimeter which exceeded three times the
amount of dysprosium measured in the same terms, by an amount of 2.97 micromoles per
cubic centimeter. Namely, there was an excess of 2.97 micromoles per cubic centimeter
of iodine to combine with all the dysprosium present in the fill material.
[0114] The lamp 7 including the fill material as described above was placed in the microwave
generating device of Fig. 1, and the light output in the wave length range of 350
to 425 nm and the starting time, i.e., the length of time that the lamp 7 required
to attain 80 percent light emission of the stable state light emission, were measured.
The light output thus measured during the stable state scored 160 in an arbitary scale,
which is 3.2 times as much as that of the lamp 7 the fill material of which includes
only mercury.
[0115] The starting time on the other hand was 22.0 seconds.
[0116] In the examples described above, examples I and III through XVIII relates to the
case in which the fill material comprises iron as the light emitting metal, and examples
XIX and XXI through XXX to the case in which the fill material comprises dysprosium
as the light emitting metal.
[0117] In the examples I and III through VI, the iron content of the fill material was changed,
and iodine was included in the fill material in the form of mercury iodide (HgI
2), in the amount substantially sufficient to combine with all the iron present in
the fill material to form iron iodide (FeI
2). Namely, when the lamp is excited by the microwave and turned on, the iodine contained
in the mercury iodide reacts with iron and forms iron iodide. When the iron iodide
thus formed ranges from 0.38 to 1.91 micromoles per cubic centimter, the light output
scored not less than 90 percent of the maximum light output attainable (example XIV)
in the case in which the fill material comtains iron as the light emitting metal.
[0118] In the examples I and VII through X, the content of iron was fixed, while that of
mercury iodide was varied so that the amount of iodine varied from the cases in which
there was a shortage of iodine to combine with all the iron present, to the cases
in which there was an axcess of iodine to combine with all the iron present. When
the excess amount of iodine in terms of atoms is not more than 2.0 micromoles per
cubic centimer, the light outputs not less than 90 percent of the maximum light output
were scored. When the amount of iodine is less than the amount thereof sufficient
to combine with all the iron present in the fill material, there remains the metallic
iron, as the amount of iron iodide formed is limited by the amount of iodine present
in the fill material. In this case, the light output is proportional to the amount
of iron iodide formed in the envelope 7a, and the inner surface of the envelope 7a
formed of quartz of the lamp 7 tends to lose transparency in less operation time thereof
than in the case in which an excess iodine is present. Thus, the preferred amount
of iodine in excess of the amount necessary to combine with all the iron present in
the fill material is from 0.02 to 0.2 micromoles per cubic centimeter of the volumetric
content of the envelope 7a.
[0119] In the examples XI through XIII, the contents of iron and mercury iodide in the fill
material were fixed, while that of mercury was changed. When the amount of mercury
in the fill material was from 17.6 to 41.3 micromoles per cubic centimeter, the light
output of the lamp 7 scored not less than 90 percent of the maximum light output attainable
in the case in which the fill material includes iron as the light emitting metal.
[0120] In the examples I and XIV through XVIII, the contents of iron, mercury iodide, and
mercury in the fill material were fixed, while the pressure of argon was changed.
When the pressure of argon was from 20 to 150 torr, the light output of the lamp 7
was not less than 95 percent of the maximum light output attainable in the case in
which the light emitting metal is iron, which maximum is attained in example XIV.
In example XIV, however, the discharge in the envelope 7a tended to be extinguished
before it reaches the stable state of light emission. It was found that the preferred
pressure of argon in the envelope 7a was from 30 to 130 torr.
[0121] In the examples XIX and XXI to XXIII; dysprosium content was varied, while iodine
in the form of mercury iodide was sealed in the envelope 7a in an amount sufficient
to combine with all the dysprosium present in the fill material. When the amount of
dysprosium was not less than 0.13 micromoles per cubic centimeter and not more than
0.39 micromoles per cubic centimeter, the lamp 7 scored not less than about 70 percent
of the maximum light output attainable in the case in which the fill material included
dysprosium as the light emitting metal.
[0122] In the examples XIX, and XXIV through XXVII, the dysprosium and the mercury iodide
contents were fixed, while the mercury content was varied. When the amount of mercury
in the fill material is from 17.6 to 53.0 micromoles per cubic centimeter, then the
lamp 7 scored not less than about 80 percent of the maximum light output attainable
in the case in which the fill material comprised dysprosium as the light emitting
metal.
[0123] In the examples XIX and XXVIII through XXX, the amount of dysprosium and the total
amount of mercury were fixed, while the excess amount of iodine was varied. These
examples also scored not less than 80 percent of the maximum light output attainable
in these cases. When the mercury iodide content was less than the amount which contains
sufficient amount of iodide to combine with all the dysprosium present in the fill
material, i.e., when an excess amount of dysprosium is present with respect to the
amount of iodine, the envelope 7a formed of quartz lost transparency thereof in a
shorter operational time than in the case in which iodine is sealed in excess with
respect to the amount of dysprosium. Thus, it is preferred that iodine is sealed in
excess of the amount necessary to combine with all the dysprosium present in the fill
material.