[0001] This invention relates to humidifiers and wore particularly to the electrode type
steam generating humidifiers.
[0002] Such humidifiers in which a current is passed-between spaced electrodes in a water-carrying
vessel are well-known. The current causes heating of the water and subsequently generation
of steam which is passed into the air space, the moisture content of which is to be
controlled.
[0003] The magnitude of the current and hence the steam generating ability is dependent
on the voltage applied to the electrodes, the size, shape and spacing of the electrodes,
the depth of immersion of the electrodes and the conductivity and volume of the water.
A serious problem which has been recognised is that the conductivity of a water supply
can vary by as much as 10:1 depending on the geological conditions of the source and
also the conductivity of a particular water supply can vary daily in the same range
(10:1) depending on variables such as interconnections in the water main and switching
and/or mixing of the sources and the season of the year. Because the design of an
evaporator is dependent on the conductivity of the water in the vessel, it is necessary
to ensure that the conductivity of the water in the vessel is maintained near the
value for which the evaporator was designed or else tedious adjustments or adaptations
of the evaporator will be necessary to accommodate different water conductivites.
[0004] Furthermore, because of the continual evaporation of water from the vessel, impurities
from the water remain in the vessel and increase in concentration to the point where
flushing of the vessel is necessary to maintain the contained water conductivity at
the designed value and to reduce rineral build-up on the electrodes. If flushing is
carried out merely on a time basis, the water may be flushed when its conductivity
is below the designed value. In any event, the loss of hot water from the system during
flushing is likely to exceed that necessary to maintain the conductivity at the designed
value.
[0005] United States Patent 3,937,920 issued on February 10, 1976 to Plascon A.G. discloses
a system which deals with these problems by arranging for the conductivity of the
water in the vessel to be maintained at a value which is considerably higher than
the average conductivity of typical water sources. As water is boiled off, the conductivity
in the vessel gradually rises to the optinum designed value.
[0006] The magnitude of the current between the electrodes is measured continuously and
the actual time taken for the current to drop between two predetermined values is
compared with the calculated time required for the current to drop between those two
values when the conductivity of the water is at the designed value. If the actual
time measured is shorter, some of the water is flushed from the vessel, as this represents
an unacceptably high conductivity.
[0007] It is an object of the present invention to provide a novel method of operating a
water-vapor generator to control the conductivity of the contained water and maintain
the conductivity at the designed value..
[0008] It is another object of the present invention to provide a novel water-vapor generator
adapted to control the conductivity of the contained water and maintain the conductivity
at the designed value.
[0009] The present invention involves obtaining a measure of the frequency of two or more
cycles, each cycle containing a boil and fill leg, and, if the measure of the frequency
increases above a predetermined value corresponding to a desired, i.e. design, frequency,
causing water to be discharged-.
[0010] The frequency can be determined in various ways, such as by measuring the time to
count a predetermined plurality of cycles, or by counting the number of cycles in
a predetermined duration.
[0011] Alternatively, the frequency of the fill legs or the boil legs can be used as a measure
of the cycle frequency in which case the time to measure a predetermined plurality
of fill legs or boil legs is obtained or the number of fill legs or boil legs in a
predetermined duration is counted.
[0012] As another alternative, the accumulated boil time over two or more cycles or the
accumulated fill time over two or more cycles may be measured and this also would
give a measure or indication of the cycle frequency.
[0013] The drain leg can be adjusted directly by time so that the higher the frequency measured
the longer the drain leg. Alternatively, the drain leg can be arranged to continue
until the current reaches- a threshold value, the higher the frequency the lower the
threshold value selected.
[0014] By concentrating on measuring the frequency of two or more cycles, the invention
has the effect of minimizing the possibility of obtaining an incorrect indication
of contained water conductivity which could arise due to spurious or transient factors
and thereby minimizing the possibility of ordering a flush cycle needlessly.
[0015] Finally, a feedback control can be built in which has the effect of altering the
requirements for a particular group of cycles depending upon the frequency measured
for the immediately preceding group of cycles. Thus, the measured frequency is compared
with a variable frequency which is weighted in relation to the frequency measured
for the preceding group. In other words, if the frequency measured for the previous
group of cycles was high, the predetermined frequency is decreased and if the frequency
measured for the previous cycles was low, the predetermined frequency is increased.
The effect of this is to stabilize the water level substantially at one point with
respect to the electrodes whether the mains supply is of high or low conductivity.
[0016] The invention will now be described in greater detail with reference to the accompanying
drawings, in which:
Figure 1 is a diagram illustrating the principles underlying the present invention;
Figure 2 is a schematic diagram illustrating an embodiment of the invention which
applies the principles illustrated in Figure 1;
Figure 3 is a diagram illustrating a modification of the basic technique illustrated
in Figure 1;
Figure 4 illustrates diagrammatically the difference in outcome between applying the
principles of Figure l,on the one hand and Figure 3 on the other hand;
Figure 5 is a diagram illustrating a variant of the basic _technique shown in Figure
1; and
Figure 6 is a diagram illustrating a further variant of the basic technique shown
in Figure 1.
[0017] It is believed that a better understanding of the circuit shown in Figure 2 will
be achieved after an explanation of the principles upon which the invention is based
is given with reference to Figure 1.
[0018] It should be appreciated that, as is well-known in the art, a vaporization container
contains two or more electrodes which heat water in the container to boiling point
and as water is boiled off and replenished with "fresh" mains water the concentration
of minerals in the container water gradually increases causing a corresponding increase
in the conductivity of the water. It is necessary to flush out some of the container
water from time to time to maintain the water conductivity around the value for which
the unit was designed. Figure 1 explains how this may be accomplished according to
the present invention.
[0019] With particular reference to Figure la which is a plot of the electrode current,
expressed as a percentage of the rated or designed value, against time, this shows
fill, boil and drain cycles of the vaporization vessel.
[0020] When the container is originally filled from an empty condition the current level
rises to a point (e.g. 5%) above the rated current at which point the water supply
is cut off. As current at 105% continues to flow the water heats to boiling point
and starts boiling off as steam to be used in air conditioning. As the water level
falls due to the loss of water as steam, the current falls until it reaches 95% of
the rated value at which point mains replenishing water is fed to the container and
the current rises once more to the 105% at which point the supply of water is cut
off and the boil/fill cycle is repeated. Because the conductivity of the water is
increasing gradually with every cycle, it takes less water to again reach 105%, and
less water needs to be boiled away to drop to 95%. Therefore, the cycling rate or
frequency of cycling increases gradually and this observation is used as the basis
of the present invention.
[0021] The point A represents the end of a fill portion of a cycle near start-up of the
whole operation, i.e. before the container water has become very concentrated. It
could, in fact, be the end of the very first fill leg beginning from empty. From the
point A the next three (for example) complete boil/fill cycles are counted and this
occurs at point B. The design frequency, i.e. the frequency which would occur if the
contained water conductivity were at the desired or designed value, can be established
theoretically or experimentally and this would give a duration C-D for three complete
cycles. If the measured frequency of the three cycles from A - B differs from the
design frequency it can be determined whether the conductivity is too high or too
low. In the case shown, the first three cycles from A - B last longer than C - D,
i.e. the measured frequency is less than the design frequency, for three complete
cycles assuming the water is at the desired conductivity. This means that the water
conductivity is less than the desired value and no drain cycle is initiated.
[0022] The end of the very next fill portion at which the 105% current level is reached
is designated E and the next three complete boil/fill cycles ending at point F are
counted and it is noted that the frequency from E - F is greater than the design frequency
indicated by C - D corresponding to the desired conductivity. In fact, the three cycles
have elapsed some 4 units before C - D and so at point F a drain cycle is initiated
to flush out some of the concentrated water. The drain cycle continues to the point
G at which the electrode current is a predetermined percentage of the rated current,
e.g. 80%. The length of the drain cycle depends on how much greater the actual frequency
is than the design frequency. If the actual frequency is very high, this indicates
that the water conductivity is very high and so the drain cycle should be correspondingly
long. The drain cycle may be arranged to last a time proportional to the difference
between the actual frequency and the design frequency or, as is preferred, the drain
cycle is arranged so that it shuts off when a predetermined value of current is reached
this value being proportional to the frequency difference. Thus, the point G at which
the drain cycle ceases corresponds to 80% current, the point H (which is the drain
shut-off point after the next three cycles) corresponds to 90% current and so on.
[0023] Figure la has been drawn for the case where the conductivity of the replenishing
water is low and Figure lb represents the case where the replenishing water is much
higher in conductivity. It can be seen that the predetermined number of cycles is
counted much more quickly, i.e. the actual frequency is much higher, and correspondingly
lower values of current are reached before the drain cycles are ceased in the case
of Figure lb.
[0024] Apparatus for carrying out the principles described above in relation to Figure 1
will now be described with reference to Figure 2 which is a schematic diagram of an
embodiment of the invention.
[0025] A vaporization container or vessel 1, constructed of a non-corrosive and non-conductive
material contains a plurality (two in the embodiment shown) of electrodes 2. The electrodes
may be concentric cylindrical tubes or spaced plates and in the embodiment shown are
spaced plates vertically disposed and of constant cross-section throughout their height.
The electrodes are constructed of a non-corrosive electrically conductive material.
[0026] The vessel 1 has an opening 3 at the top thereof-for communication with a steam-carrying
conduit (not shown) for conveying steam to a space the moisture content or humidity
of which is to be controlled.
[0027] The vessel I has an opening at its bottom, the opening communicating by means of
a pipe 5 with a drain valve (solenoid operated in this embodiment) 6. A branch pipe
7 connects opening 4 to a solenoid operated fill valve 8 for supplying replenishing
mains water to the vessel.
[0028] The electrodes 2 are connected to an electrical supply source 8 by means of lead
wires 9a and 9b. In line 9b a current sensing means, typically a current transformer
10 or an electrical resistor, is provided. The transformer outputs are connected by
lead wires lla and llb to a transducer 12 which processes the signals from the current
transformer into control signals proportional to the electrode current. The current
transformer and transducer together form a measuring device for the current passing
between the electrodes. A manual or automatically adjusting variable resistor 13 is
provided in line lla so as to permit regulation of the magnitude of the control signal.
[0029] The output of the transducer 12 is connected via lead 14 to a fill threshold switch
15 which contains a changeover switch having three contacts 15a, 15b and 15c and a
movable contact 15a which is movable from the position shown in the solid line to
the position shown in the broken line. The threshold switch 15 is constructed so that
the movable contact 15d moves to the position shown in solid when the control signal
reaches or exceeds a predetermined maximum value fixed in the threshold switch and
switches back to the broken line position when the control signal drops to or below
a predcternined minimum value fixed in the threshold switch. The value of the control
signal is determined by the setting of variable resistor 13 and corresponds to the
value of the current which should flow through the electrodes 2 and thus is also a
measure of the value of the vaporizing capacity. Since the relationship between the
electrode current and the magnitude of the control signal can be set by adjusting
the variable resistor 13, the response values of the threshold switch 15 can be adjusted
over a wide capacity range of the vaporizer. Whatever the actual current values chosen,
typically the maximum and minimum values at which the threshold switches cause changeover
of contact 15d represent, respectively, 105% and 95% of the rated or designed current
value.
[0030] The output of transducer 12 is also connected through lead wire 16, an extension
of lead 14, to drain thresholds 17a - 17e which are, respectively, fixed to turn off
when the control signal drops below a value corresponding to 95%, 90%, 85%, 80%, 75%
of the rated electrode current. Of course, additional drain thresholds going down
to 40% or so, as necessary, would also be incorporated in the circuit but it is unnecessary
to show all of these.
[0031] A control voltage source 20 is connected via a line 21,22 to contact 15a of the changeover
switch. Contact 15b is connected to a cycle counter 23 via line 24 which branches
via line 25 to a' start activator 26 which is also connected directly via line 21
to voltage source 20. The start actuator output is connected to the start input of
a count down sequencer 27 via lines 28 and 29 and to the reset input of cycle counter
23 via lines 28 and 30. The cycle counter 23 has an output which is connected via
line 32 to a series of switches 33a-33f of which switches 33b-33f correspond, respectively
to drain thresholds 17a-17e and are shown connected via lines 34.
[0032] An output line 35 from switch 33a is connected back to the start input of sequencer
27 and the reset input of counter 23.
[0033] Although switches 33 and thresholds 17 are shown separately, they may be combined
as relays each when energised via line 32 passing a drain opening signal when the
predetermined threshold is reached. Accordingly, hereinafter switches 33 will be referred
to as relays and it will be understood that thresholds 17 are not separate physical
elements but represent preset thresholds of the relays.
[0034] A line 36 is connected from line 21 to the relays 33 (shown connected to drain thresholds
17) and an output line 37 is connected from the relays 22 (shown connected to the
drain thresholds 17a-17e) to a relay 38 which has a normally open contact arranged
38a in the line 37 between the drain thresholds and the solenoid drain valve 6. Relay
38 also has a normally closed contact 38b arranged in a line 40 which connects contact
15c of the changeover switch to solenoid fill valve 8.
[0035] The count down sequencer 27 is arranged to step through relays 33n to 33a in that
order as it counts down from the preset value, for example 20.
[0036] The circuit of Figure 2 operates as follows. Initially solenoid fill valve 8 is open
to fill or partially fill the vessel 1. When the current passing between electrodes
2 reaches the 105% value as monitored by the current measuring device (current transformer
10 and transducer 13) the threshold switch 15 causes the changeover switch to move
to the position shown in full from the broken line position thus immediately closing
the fill valve.
[0037] The switching over of the changeover switch causes line 24 and cycle counter 23 to
be energized by voltage source 20 and this is seen by cycle counter 23 as the start
of a cycle. At the same time the count down sequencer 27 is started and begins to
step through the relays 33.
[0038] When the current value reaches 95% due to boiling off of some of the water in vessel
1 the fill threshold switch 15 changes the changeover switch back to the broken line
position and fill valve 8 opens again to replenish the vessel with mains water. Again
the 105% value is reached and the changeover switch reverts to the solid line position
signalling to cycle counter 23 one complete cycle.
[0039] This is repeated until the cycle counter 23 has counted a predetermined number of
cycles set in the cycle counter at which point an output signal is generated by the
cycle counter 23 and appears on line 32.
[0040] If the count down sequencer has previously completed counting to zero the relays
33 will have been stepped down to 33a and the cycle counter 23 will simply be switched
through relay 33a to line 35 to reset the cycle counter 23 and restart the count down
timer 27. This corresponds to point B on Figure la.
[0041] If the count down sequencer 27 has not run out but is currently switched on to one
of relays 33b-33n at the time the cycle counter output signal is derived, this signal
energises that relay which causes a signal to pass from voltage source 20 to relay
38. This corresponds to point F on Figure la. For example, if relay 33e is currently
opened by count down sequencer 27, the cycle counter output signal energises this
relay (thus energising from voltage source 20 relay 38) until the current value drops
to 80%. It can be seen from Figure la that this current value corresponds to the step
(4) to which the sequencer has progressed. The energising of relay 38 causes contact
38a to close thus activating drain valve 6 and causing water to drain from vessel
1. At the same time contact 38b is opened thus presenting an energising signal to
reach solenoid fill valve 8 which would occur when the current dropped to the 95%
value set in the threshold switch 15. Thus, the fill valve is prevented from being
open while the drain valve is open. In certain cases it may be considered advantageous
to allow replenishing water to be mixed in while the drain valve is open and to achieve
that, relay 38 could be dispensed with so that when the current value dropped to 95%
the fill threshold switch 15 would energise the fill valve 8. Alternatively, relay
38 could be adapted by replacing normally closed contact 38b with a normally open
contact connected directly between voltage source 20 and fill solenoid valve 8 so
that as soon as drain valve 6 was opened the fill valve would also open.
[0042] When the current passing between electrodes 2 drops to 80% threshold 17d cuts off
drain valve and recloses contact 38b allowing fill valve 8 to open. This corresponds
to point G on Figure la. When the vessel is refilled the same sequence of steps is
repeated. The cycle counter 23 is reset and count down sequencer 27 restarted when
the changeover switch moves once more to the full line position.
[0043] Figure 3 illustrates a modification of the basic techniques described with reference
to Figure I. Here the value to which the count down sequencer 27 is reset depends
upon the duration of the previous three cycles. Figure 3a illustrates the case where
the mains water is of low conductivity and Figure 3b where the mains water is of high
conductivity. Looking at the first three cycles of Figure 3a, it can be seen that
2 units remain on the count down sequencer after three cycles have been connected.
As before the drain cycle is carried out over portion M-N until a threshold current
value corresponding to the step to which the sequencer has progressed is reached.
At the end of the fill cycle N-0 the count down sequencer is reset by adding a predetermined
number of steps to the steps left on the sequencer at point M. Thus, for the next
three cycles, i.e. 0-P the count down sequencer counts down from 2 + 15 = 17.
[0044] When the point P is reached there are still 3 steps remaining in the count down sequencer
and the drain cycle P-Q corresponds to this value. Also, 15 steps are added to this
value to provide the new reset quantity of 18 steps to the count down sequencer for
the next three cycles and so on.
[0045] As can be seen by comparing Figure 3a and 3b, when the water conductivity is very
high and cycling is correspondingly fast, the drain cycles are longer as was the case
with the basic system of Figures la and lb. As can be seen by comparing Figures 3b
and lb the drain cycles are longer in Figure 3b due to the weighting of the reset
quantity of the count down sequencer. The effect of this weighting in practical terms
can be appreciated by comparing Figures 4a and 4b.
[0046] Figure 4a illustrates diagrammatically the difference in water levels inside the
vessel for mains water of high conductivity (I) and for mains water of low conductivity
(II) using a fixed number of count down steps as explained with reference to Figure
1 and Figures 4b(I) and 4b(II) illustrate the same thing but using a "weighted" number
of count down steps. In Figures 4a(I) and 4a(II) reference numeral 40 indicates the
water level in the high conductivity case at which the current is 100% of designed
value and the same current value is achieved at water level 41 when the conductivity
is low. It can be seen that there is a considerable difference in these levels. Because
the water level varies considerably over the range of water conductivities which would
be expected from different water supplies it is not possible to design the vessel
so that when it is new the water level for a predetermined current value is established
just above the lower ends of the electrodes 2 for all values of water conductivity.
[0047] With reference now to Figure 4b, it can be seen that by "weighting" the number of
count down steps, the levels 40' and 41', representing the high conductivity and low
conductivity levels, respectively, for 100% current are virtually identical and so
the vessel can be designed so that this level is near the bottom of the electrodes
2. This is desirable because in this way, the bottom of the electrodes is used first
and as this becomes encrusted with deposits the level moves up bringing into use fresh
surface portions of the electrodes. Moreover, a low water level means that the current
is passed between a small surface area of the electrodes and the relatively high current
density enables great penetration of the build-up of deposits on the electrodes. Thus,
the deposits can build-up to a considerable thickness before the current flow is seriously
curtailed which means a longer cylinder life.
[0048] In order to achieve the "weighting" of the count down time in practice a reset corrector
45 shown in phantom in Figure 2 is provided: This has an input connected to relays
33 and an output connected to count down sequencer 27.Reset corrector operates to
add to a fixed value (e.g. 15 minutes) in the count down timer 27 the value in minutes
corresponding to the relay 33 which has previously been stepped by the count down
timer 27 while the cycle counter gave out its output signal. For example, if relay
33e were energised, the correction value may be 4 minutes so that the count down timer
would be reset at 4 + 15 = 19 minutes.
[0049] Figure 5 illustrates a variant of the technique shown in Figure 1. Here, instead
of counting up to a predetermined number of cycles, e.g. 3, and relating to whether
or not the predetermined time has expired, the total number of cycles in a predetermined
time interval is counted and compared to the number of cycles that would occur in
that same time if contained water was at design conductivity. Thus in the first interval
Tg, four cycles are counted and this number has been previously determined as the
correct number in the particular time period to give the required conductivity level.
As the conductivity increases, five cycles are counted in the next interval Tg and
a drain cycle is initiated, the extent of the drain cycle depending on the number
of counted cycles by which the predetcrnined number, namely 4, is exceeded. In the
third interval Tg shown, the number of cycles counted is 6 and the drain cycle is
correspondingly longer. Again, the length of the drain cycle can be made time dependent
or can be determined by a current threshold.
[0050] It can be seen that the technique illustrated in Figure 5 is another way of comparing
the actual frequency with the design frequency of the cycles.
[0051] Figure 6 illustrates a variant of the basic technique shown in Figure 3. Here, instead
of measuring the frequency of complete cycles, we measure the frequency of occurence
of a particular leg of the cycles, e.g. the boil leg or the fill leg. Thus, the boil
legs RS, R'S', R"S" etc. or fill legs TR, SR', S'R" etc. are counted in the predetermined
duration Tg. As the number of boil legs (or fill legs) is the same as the number of
cycles, again we are comparing the actual frequency of the cycles with the design
frequency.
[0052] Of course, instead of counting the number of boil legs or fill legs in a predetermined
time we could, in a manner analagous to the technique shown in Figure 1, keep the
number of cycles (i.e. number of boil legs or fill legs) constant and relate this
to whether or not the predetermined duration has expired. This, again amounts to comparing
the actual frequency of the cycles with the design frequency.
[0053] As another variation of the basic technique, instead of counting the number of boil
legs or fill legs in a predetermined duration Tg, the actual accumulated duration
of all the boil legs (or fill legs as the case may be) could be obtained as a measure
of the cycle frequency (The accumulated time is, of course, inversely proportional
to the frequency). This then would be compared with a value corresponding to the accumulated
boil (or fill) time which would be expected for a design frequency.
[0054] It will be understood that this invention is capable of a variety of modifications
and variations which will become apparent to those skilled in the art upon a reading
of the specification, such modifications intended to be part of the invention as defined
in the appended claims. For example, it should be apparent that the apparatus shown
in Figure 2 is merely exemplary and that the invention could be realised by electronic
or fully integrated micropressor circuitry instead of the electromechanical circuitry
shown.
1. A method of operating a water-vapor generator comprising a vaporization vessel
containing water and having electrodes which are connected to a power supply between
which flows a current the magnitude of which depends on the depth of immersion of
the electrodes in the water in the vessel and the conductivity of the water, the method
comprising continuously measuring the magnitude of the electrode current, filling
the vessel with water to an extent to give a predetermined maximum electrode current,
allowing water in the vessel to boil off until a predetermined minimum electrode current
is achieved, re-filling the vessel to obtain again the predetermined maximum electrode
current and repeating the boiling and filling steps to provide a plurality of cycles
each containing a boil leg and a fill leg, obtaining a measure of the frequency of
each of successive pluralities of cycles and comparing the measure obtained with a
predetermined frequency value corresponding to a desired conductivity of the water,
and if the measured frequency is greater than the predetermined frequency, automatically
discharging from the vessel a quantity of water dependent upon the amount by which
the measured frequency exceeds the predetermined frequency.
2. A method according to clain 1, in which the measure of the frequency is obtained
by counting the cycles and measuring the time taken to count a predetermined number
of cycles, a quantity of water being discharged if the time taken is less than a predetermined
time for the predetermined number of cycles.
3. A method according to claim 1, in which the water is discharged from the vessel
until a selected one of a predeternined plurality of low current thresholds is achieved
by the current passing between the electrodes, the greater the measured frequency
exceeds the.predctermined frequency the lower the particular low current threshold
selected.
4. A method according to claim 2 in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing between the electrodes, the shorter the time taken to count
the predetermined number of cycles the lower the low current threshold selected.
5. A method according to claim 2, including varying the predetermined time for the
predetermined number of cycles, the predetermined time for a particular group of cycles
being increased above a fixed value in inverse proportion to the time measured for
the immediately preceding predetermined number of cycles.
6. A method according to claim 5, in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing between the electrodes, the greater the amount by which the
time taken to count the predetermined number of cycles is exceeded by the predetermined
time in effect for that particular predetermined number of cycles the lower the low
current threshold selected.
7. A method according to claim 1, in which the measure of the frequency is obtained
by counting the cycles completed in a predetermined duration, a quantity of water
being discharged if the number of cycles 1 counted in the predetermined duration is greater than a predetermined number.
8. A method according to claim 7, in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing between the electrodes, the greater the number of cycles counted
the lower the low current threshold selected.
9. A method according to claim 1, in which the measure of the frequency is obtained
by measuring the time of the boil legs of the cycles and adding the total boil time
over a predetermined interval, a quantity of water being discharged if the total boil
time is less than a predetermined total accumulated boil time for the predetermined
interval.
10. A method according to claim 9 in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the less the total boil time measured
the lower the low current threshold selected.
11. A method according to claim 1, in which the measure of the frequency is obtained
by measuring the time of the fill legs of the cycles and adding the total fill time
over a predetermined interval, a quantity of water being discharged if the total fill
time is less than a predetermined accumulated total fill time for the predetermined
interval.
12. A method according to claim 11 in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the less the total fill time measured
the lower the low current threshold selected.
13. A method according to claim 1, in which the measure of the frequency is obtained
by counting the number of boil legs completed in a predetermined duration, a quantity
of water being discharged if the number of boil legs counted in the predetermined
duration is greater than a predetermined number.
14. A method according to claim 13, in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the greater the number of boil legs
counted the lower the low current threshold selected.
15. A method according to claim 1, in which the measure of the frequency is obtained
by counting the boil legs and measuring the time . taken to count a predetermined
number of boil legs, a quantity of water being discharged if the time taken is less
than a predetermined time for the predetermined number of cycles.
16. A method according to claim 15 in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the shorter the time taken to count
the predetermined number of boil legs the lower the low current threshold selected.
17. A method according to claim 1, in which the measure of the frequency is obtained
by counting the number of fill legs completed in a predetermined duration, a quantity
of water being discharged if the number of fill legs counted in the predetermined
duration is greater than a predetermined number.
18. A method according to claim 17, in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the greater the number of fill legs
counted the lower the low current threshold selected.
19. A method according to claim 1, in which the measure of the frequency is obtained
by counting the fill legs and measuring the time taken to count a predetermined number
of fill legs, a quantity of water being discharged if the time taken is less than
a predetermined time for the predetermined number of cycles.
20. A method according to claim 19 in which the water is discharged from the vessel
until a selected one of a predetermined plurality of low current thresholds is achieved
by the current passing through the electrodes, the shorter the tire taken to count
the predetermined number of fill legs the lower the low current threshold selected.
21. A water-vapor generator comprising a vaporization vessel provided with electrodes
and with an inlet for fresh water and an outlet for discharging water in order to
reduce the concentration of minerals an inlet valve arranged to control the flow of
water through the inlet, an outlet valve arranged to control the flow of water through
the outlet, means for measuring continuously the current flowing through the electrodes,
threshold means connected to the current measuring means and operable to open the
inlet valve when the measured current reaches a predetermined minimum value and to
close the inlet valve when the measured current reaches a predetermined maximum value,
means connected to the threshold means for obtaining a measure of the frequency of
successive pluralities of cycles each containing a leg descending from the predetermined
maximum value to the predetermined minimum value and the subsequent leg ascending
from the predetermined minimum value to the predetermined maximum value, means for
comparing the frequency measure obtained with a predetermined frequency value corresponding
to a desired conductivity of the water and a control device connected to be controlled
by the comparison means to open the outlet valve when the measured frequency is greater
than the predetermined frequency and cause the outlet valve to remain open for a time
interval proportional to the amount by which the measured frequency exceeds the predetermined
frequency.
22. A water-vapor generator comprising a vaporization vessel provided with electrodes
and with an inlet for fresh water and an outlet for discharging water in order to
reduce the concentration of minerals, an inlet valve arranged to control the flow
of water through the inlet, an outlet valve arranged to control the flow of water
through the outlet, means for measuring continuously the current flowing between the
electrodes threshold means connected to the current measuring means and operable to
open the inlet valve when the measured current reaches a predetermined minimum value
and to close the inlet valve when the measured current reaches a predetermined maximum
value, means connected to the threshold means for counting successive pluralities
of cycles each containing a leg descending from the predetermined maximum value to
the predetermined minimum value and the subsequent leg ascending from the predetermined
minimum value to the predetermined maximum value, means for measuring the duration
of a predetermined number of cycles and for comparing the duration with a predetermined
duration corresponding to a desired conductivity of the water and a control device
connected to the outlet valve and operable under control of the measuring and comparing
means to open the outlet valve when the actual duration is less than the predetermined
duration and cause the outlet valve to remain open for a time interval proportional
to the amount by which the predetermined duration exceeds the actual duration.
23. A water-vapor generator according to claim 22 wherein the measuring and comparing
means comprises a count down sequencer in which the predetermined duration is set,
the count down sequencer being arranged to step down through a plurality of switch
means connected to the control device and wherein the means for counting cycles has
an output connected to the plurality of switch means, an energising signal being obtained
at the output of the means for counting cycles when the predetermined number of cycles
has been counted whereby the energising signal passes through one of the plurality
of switch means to the control device when the count down sequencer has not finished
counting down to the predetermined duration.
24. A water-vapor generator according to claim 23 wherein the plurality of switch
means comprises a plurality of relays connected to the means for measuring continuously
the current flowing through the electrodes, the relays having ascending energising
threshold values in direct relation to the order in which the count down sequencer
counts down through the relays, whereby a relay energised will remain energised until
the electrode current drops below the respective threshold value.
25. A water-vapor generator according to claim 24 including a reset corrector connected
to the relays and to a reset input of the count down sequencer, the reset corrector
being operable to increase the predetermined duration set into the count down sequencer
after the count down scquencer has counted out, the amount by which the reset corrector
increases the predetermined duration depending upon which of the pluralities of relays
was immediately previously energised, the amount decreasing in direct relation to
the order in which the count down sequencer counts through the relays.
26. A water-vapor generator comprising a vaporization vessel provided with electrodes
and with an inlet for fresh water and an outlet for discharging water in order to
reduce the concentration of minerals an inlet valve arranged to control the flow of
water through the inlet, an outlet valve arranged to control the flow of-water through
the outlet, means for measuring continuously the current flowing through the electrodes,
threshold means connected to the current measuring means and operable to open the
inlet valve when the measured current reaches a predetermined minimum value and to
close the inlet valve when the measured current reaches a predetermined maximum value,
means for counting successively the number of cycles during a predetermined duration,
each cycle containing a leg descending from the predetermined maximum value and a
leg ascending from the predetermined minimum value, means for comparing the number
of cycles counted with a predetermined number of cycles corresponding to a desired
conductivity of the water and a control device connected to the outlet valve and operable
under control of the means for comparing to open the outlet valve when the number
counted -is greater than the predetermined number and cause the outlet valve to remain
open for a duration proportional to the amount by which the number counted exceeds
the predetermined number.