[0001] The invention relates to a method and a device to store and display a history of
Kt/V or other adequacy parameters trends in a machine able to perform kidney substitution
treatments. The stored trends can be evaluated against the ongoing treatment offering
some parameters, which may be used by the medical staff to assess said ongoing treatment.
[0002] Dialysis adequacy is the topic that has got and gets more attention when one thinks
about patient outcome. In order to estimate dialysis adequacy one needs a parameter
establishing a relation between dialysis dosage and patient outcome. The most accepted
parameter to estimate the quantity of dialysis delivered or dosage is the Kt/V, where
K is the effective clearance for urea, t is the treatment time and V is the urea distribution
volume which matches the total body water.
[0003] The NCDS (National Cooperative Dialysis Study) and the HEMO study found, after analyzing
a large patient group, that morbidity and mortality in end stage renal disease (ESRD)
was strongly correlated with the Kt/V value or dialysis dose. Data obtained from these
studies resulted in guidelines regarding hemodialysis treatments, which demand a minimum
dose of Kt/V-1.2 for non-diabetic patients and 1.4 for diabetics (DOQ) guidclines).
It is worthy to point out that a morbidity decrease not only improves the patient
well-being, but also reduces Significantly the medical costs as the patient requires
less care.
[0004] The need of a reliable and cost effective method to monitor the Kt/V and by extension
control dialysis adequacy and morbidity, is therefore easily understood.
[0005] In the Kt/V calculation, the main problems are K and V estimation along with the
multi-compartment urea kinetics. V can be estimated by bioimpedance, anthropometric
measurements or applying the urea kinetic model (UKM), all these methods have a certain
degree of error. K can be estimated so far by measuring the urea blood concentration
before and after the treatment or by monitoring inlet and outlet conductivity changes
in the dialysate side.
[0006] 13lood samples method is the reference one. After taking the blood samples and applying
either UKM or Daugirdas formula a single pool Kt/V (spKt/V) is estimated, further,
[0007] Daugirdas second generation formulas should be used to get an equilibrated Kt/V (eKt/V)
which accounts for the urea rebound caused by the fact that urea kinetic's does not
follow a single pool model but a multi-compartment one. This method has two main problems:
it is not possible to know whether the treatment is adequate or not before it finishes,
therefore it is not possible to perform any action to improve the situation; it is
not an easy to apply method: sampling time is very important to get an accurate value,
and the medical stuff must send the samples to the lab, wait for the results and calculate
Kt/V values with the help of a computer. These facts result on a monthly basis Kt/V
measurements in best case, which means that in worst case scenario a patient might
be under-dialyzed for one whole month.
[0008] Conductivity methods are based on the observation that sodium clearance is almost
equal to urea clearance and that the relationship between dialysate conductivity and
dialysate sodium concentration can be considered linear on the temperature range of
interest. Therefore it is possible to get urea clearance by measuring the sodium diffusion
transport through the membrane in the dialyzer.
[0009] It is important to introduce the concept of Dialysance, as it slightly differs from
Clearance:
Clearance is defined as the ratio between transport rate and concentration multiplied
by flow, and it is applicable when the diffusing substance is on the blood side but
not on the dialysate, that is the case for urea.
[0010] Dialysance is defined as the ratio between transport rate and concentration gradient
multiplied by flow, and it is applicable when the diffusing substance is in both dialyzer
sides. When one applies conductivity methods to measure urea Clearance, one actually
measures sodium Dialysance (Depner T, Garred I.. Solute transport mechanisms in dialysis.
Hörl W, Koch K, Lindsay R, Ronco C, Winchester JF, editors. Replacement of renal function
by dialysis, 5
th ed. Kluwer academic publishers, 2004:73-91).
[0011] During conductivity based clearance measurements, a dialysate inlet conductivity
different to the blood one is produced, which results in a net transfer of sodium
either from blood to dialysate or from dialysate to blood due to the generated gradient.
There are currently several methods which are applied in the industry:
In a first method a one-step conductivity profile is performed; in a second method
a two-step conductivity profile is performed; and in a third method an integration
of conductivity peaks is used. (Polaschegg HD, Levin NW. Hemodialysis machines and monitoris. Hörl W, Koch K, Lindsay
R, Ronco C, Winchester JF, editors. Replacement of renal function by dialysis, 5th
ed. Kluwer academic publishers, 2004:414-418). The main advantage of this approach is that it is relatively easy to implement
and cost effective as it only needs an extra conductivity/temperature sensor downstream
the dialyzer. It offers Kt/V measurements during the treatment allowing the medical
staff to react and perform some actions in case the treatment is not going as it should.
However, conductivity based methods have also some limitations: they can induce some
sodium load in the patient during the measurement; they are not useful to obtain other
interesting parameters like nPCR. or TRU. The maximum measurement frequency offered
so far by the industry is about 20 minutes, which means that in worst case scenario
the patient could be under-dialyzed for 20 minutes. And although there are some publications
claiming it, so far, conductivity methods haven't been applied with enough reliability
to hemofiltration or hemodiafiltration treatments.
[0012] Another method to estimate hemodialysis adequacy is by direct measurement of the
waste products (urea) concentration in the effluent dialysate, this method assumes
that the evolution of urea concentration over the time in the dialysate side is proportional
to the one in the blood, therefore the slope of the line obtained after applying the
natural logarithm to the registered concentration values over the time will be the
same on both sides: dialysate side and blood side. And by definition such slope is
K/V, which multiplied by the therapy time results in the Kt/V value.
[0013] There arc two different methods available to measure online the concentration of
waste products in effluent dialysate: Urea sensors and UV spectrophotometry.
[0015] By direct measuring waste products in spent dialysate, a curve representing the concentration
or absorbance decay of uremic compounds over the whole dialysis period is obtained.
These curves sightly reflect the course of the treatment and manifest patient based
haemodynamic singularities, which influence the dialysis efficiency. The obtained
concentration-curves could be saved either in the dialysis machine or in a patient
card. During a dialysis procedure, the saved curves could be displayed by user request.
Visual comparisons between the ongoing and the previously recorded therapies could
help the physician to identify problems Moreover, a comparison algorithm could generate
some parameters to quantify these discrepancies. A signal could be triggered when
the differences get over a certain threshold to draw the attention of the medical
staff, and help them on deciding if an action is necessary.
[0016] The problem of this invention is to provide a reliable method and device to record
and display concentration or absorbance curves obtained during different dialysis
therapies; and to offer objective parameters evaluating the differences between the
ongoing and previous therapies. This problem is solved by a method with the features
described in claim 1. Furthermore this problem is solved by a device with the features
described in claim 11, Preferred embodiments of the invention are described in the
claims 2 to 10.
[0017] Further goals, advantages, features and possibilities of use of this invention arise
out of the subsequent description of the embodiments of the invention. Therefore every
described or depict feature of its own or in arbitrary meaningful combination forms
the subject matter of the invention even independent of its summary in the claims
or its reference to other claims.
[0018] It shows:
- FIG 1
- Depicts a portion of a conventional dialysis machine plus a slight modification to
host a sensor coupled with the dialysate circuit,
- FIG 2
- Shows an ongoing therapy concentration curve (solid line) together with a recorded
concentration curve (dashed line).
- FIG 3
- Shows an ongoing therapy Kt/V curve (solid line) together with a recorded concentration
curve (dashed line).
- FIG 4
- Shows a dialysis machine screen with a menu, which allows the user to select a recorded
KtIV curve.
[0019] FIG. 1 shows a draw of the dialysate circuit of a conventional dialysis machine plus
a slight modification to host a sensor coupled with the dialysate circuit. The blood
from a patient is taken out into an extracorporcal circuit, it flows through the tube
32 into the blood chamber 30 of a dialyzer and returns to the patient through the
tube 31. The flow rate of the blood circuit is controlled by the blood pump 33. The
dialysis fluid is made of several concentrates and water, therefore the machine disclosed
in figure 1 comprises a water inlet 12, two concentrates inlets 16 and 18 and two
concentrate pumps 17 and 19.
[0020] The water flow together with the concentrates flow defines the final properties of
the dialysis fluid. The conduit 20 takes the dialysis fluid to the dialysate chamber
29 of the dialyzer, which is separated from the blood chamber 30 by a semi permeable
membrane.
[0021] The dialysis fluid it is pumped into the dialyzer by the pump 21. A second pump 34
sucks the dialysis fluid and any ultrafiltrate removed from the blood. A bypass line
35 is arranged between the pumps 21 and 34. Several valves 26, 27 and 28 are arranged
to control the dialysate flow. The conduit 36 leads the spent dialysate to a UV-sensor
37 measuring its light absorbance, the UV-sensor 37 is connected by an interface with
the computer 14 which processes the measured data, the result of the data processing
is displayed and/or printed by the device 15, which is connected with the computer
14 by an interface. The conduit 36 leads the spent dialysate after its measurement
by the U V-sensor 37 to the drain system 13. The dotted lines 22, 24 and 25 represent
an adaptation of the disclosed apparatus for hemodiafiltration treatments. The substitution
fluid comes from a substitution fluid source 11, flows through the line 22 and is
pumped in the blood lines of the patient by the pump 23. In case of post dilution
hemodiafiltration the conduit 24 leads the substitution fluid to the venous line of
the extracorporeal blood system; in case of pre dilution hemodiafiltration the conduit
25 leads the substitution fluid to the arterial line of the extracorporeal blood system;
and in case of pre-post dilution hemodiafiltration both conduits 24 and 25 are used.
The computer 14 controls all the elements shown on the figure by means of proper interfaces,
said interfaces arc not drawn for the sake of simplicity. The computer 14 gathers
information about other parameters of the dialysis machine, like for example blood
flow, dialysate flow and/or therapy time, these parameters together with the measured
data are processed, the result tunes the Kt/V measuring functionality to assess deviations.
[0022] The UV-sensor 37 can be substituted by an Urea-sensor, in this case will the urea
concentration in spent dialysate measured instead of the light absorbance. The disclosed
dialysis machine is provided with several other means as is conventional. These other
means arc not disclosed, since they are not relevant for the operation of the present
invention.
[0023] The obtained absorbance curves are stored in a patient card attached to the computer
14 or in a database hosted in the computer 14. The number of curves which may be saved
is variable and depends on the capacity of the storage medium. In our preferred embodiment
the last 15 treatments are stored in a suitable patient card. The stored treatments
are overwritten following a FIFO (first in first out) procedure. Additionally, a given
therapy can be set as not over-writable, in such a case the therapy will be kept until
it is set back to over-writable. It is also possible to store Kt/V or URR curves.
[0024] The different stored curves can be displayed by selecting them on a menu in the display
of the dialysis machine (Fig. 4). In the preferred embodiment each curve is identified
by the weekday and date when the therapy took place. The recorded curve is depicted
below the currently ongoing curve with a lighter colour. The trend of the ongoing
curve can be then visually compared with the previously recorded curve. One or more
recorded curves can be displayed simultancously.
[0025] Specific curves can be identified by special names. In our preferred embodiment the
treatment curve which delivered the best kt/V is saved as "Best treatment"; and the
treatment which delivered the worst Kt/V is saved as "Worst treatment". These specific
treatments can be kept out of the FIFO overwriting functionality, being only respectively
overwritten if a better or worse treatment is recorded. The tracking of specific treatments
may be reset by user action or after a predefined amount of time.
[0026] In our preferred embodiment an average curve out of all the curves stored in the
patient card is calculate and identified as "average". This average curve is more
valuable when comparing it against the ongoing curve because is less affected by extreme
values and therefore more representative of the patient status.
[0027] The main factors affecting the shape of the absorbance curves are the blood flow
and the dialysate flow. The stronger the counter current effect is, the more efficient
the dialysis treatment is, and by extension the more steeper the UV-absorbance curve
is. Ideally the recorded curves should be normalized to the ongoing treatment parameters
to make them comparable.
[0029] The mass balance of any compound during a dialysis therapy can be described as follows:

where K is the dialyzer clearance; C
b is The blood concentration; Q
d is the dialysate flow, and C
d is the dialysate concentration.
[0030] Because our goal is to normalize recorded treatments to the current treatment parameters
regardless of the blood concentration, we can consider C
b, as a constant. Thus, it is possible to normalize each point of a recorded concentration
curve to the ongoing curve as follows:

Where C
d-current is the normalized dialysate concentration; or in other words, the concentration that
we would get in the recorded treatment if we had use the blood flow and dialysate
flow of the current treatment.
[0031] By knowing the PS of the dialyzer it is possible to use equation I to calculate the
K's, which are required by equation 4. It is also possible to use UV-absorbance values
instead of concentrations in all the equations disclosed above.
[0032] In the preferred embodiment the PS factors four each dialyzer, type are available
in the dialysis machine, new PS factors for other dialyzer types can be added at any
time either by the technician servicing the machine or by the medical staff. The recorded
therapy data includes the dialysate absorbance, blood flow and dialysate flow at each
time point, and the PS factor. Before the start of the treatment the user selects
the dialyzer. With the dialyzer selection and the current Q
d, the recorded curve is normalized to the current curve on real-time. If for example
during the first half of a four hours therapy, the patient is treated with a K of
240 ml/min (300 ml/min blood flow) and during the second half with a K of 180 ml/min
(200 ml/min blood flow), the recorded curve will be normalize to 240 ml/min for 2
hours and to 180 ml/min for 2 hours.
[0033] The normalization of the recorded curves to the current one is not mandatory. The
dialysis machine offers the user the possibility to switch between two modes: normalized
and not normalized.
[0034] During a dialysis procedure the dialysate flow prescription may change, such a change
has important influence on the plotted concentration due to dilution effeets. An increase
of dialysate flow from 400 ml/min to 800 ml/min will cause a 50% drop of the measured
concentration, which can be misleading. It is important to account for dilution effects
during a treatment. In our preferred embodiment the plotted concentration is on real
time normalized to a standard dialysate flow of 500 ml/min. The normalization is achieved
with the following equation:

Where C
n is normalized concentration; C is concentration; Q
d-std is the standard dialysate flow to which the concentration is normalized; and Q
d-current is the current dialysate flow. The concentrations may be substituted by UV-absorbance
values.
[0035] The dilution effects can be illustrated by the following example. Let's suppose that
the treatment is started with a dialysate flow of 500 ml/min and the first absorbance
value is 3; after one hour the flow is set to 400 and the obtained absorbance is 3.1;
furthermore, after two hours the flow is set to 800 ml/min and the obtained absorbance
is 1.3. The normalized absorbance at each of the time points will be:

This approach avoids misleading the medical staff.
[0036] A visual comparison between recorded and ongoing treatments supposes a valuable help
for the experience physician, however an objective parameter that evaluates the difference
between the compared therapies would help on taking objective decisions regarding
the current treatment or future prescriptions. In the preferred embodiment the difference
between the area under the curve between the recorded and the current therapy is calculated.
The integration of the concentration curve may be or not multiplied by the dialysate
flow, in any of the cases it gives an idea of the amount of substances that were cleared
from the patient. The difference between the two therapies quantifies how much more
(or less) substances were removed.
1. A kidney substitution treatment machine for normalizing, displaying and storing curves
describing the efficiency of a kidney substitution treatment or concentrations during
said kidney substitution treatment,
with an extracorporeal blood system,
with a dialyzer, which is divided by a semi-permeable membrane into a blood chamber
and a dialyzing fluid chamber
with means to adjust a flow rate of the blood at a preset blood flow rate to pump
the patient blood through a blood chamber of the dialyzer,
with means to adjust a flow rate of a dialyzing fluid at a preset flow rate through
the dialyzing fluid system of the machine wherein the dialyzing fluid collects the
waste products from the patient after flowing through the dialyzing fluid chamber
of the dialyzer and
with means to measure continuously any kidney substitution treatment related waste
product to deliver together with the data provided by the kidney substitution treatment
machine an adequacy parameter wherein the means are coupled with the dialyzing fluid
system of the kidney substitution treatment machine and
with means to generate curves or functions of the adequacy parameter with means to
store curves or functions of the adequacy parameter with means to display curves or
functions of the adequacy parameter wherein the curves of the adequacy parameter of
the kidney substitution treatment arc normalized to make them comparable, and/or stored
in an adequate media and/or displayed on a user interface of the machine delivering
the kidney substitution treatment.
2. The kidney substitution treatment machine according to claim 1, with means to calculate
parameters quantifying the adequacy differences between treatments.
3. The kidney substitution treatment machine according to claim 1 or 2, wherein the kidney
substitution treatment is double needle hemodialysis, single needle hemodialysis,
single needle cross over hemodialysis, post-dilution hemodiafiltration, pre-dilution
hemodiafiltration, pre-post-dilution hemodiafiltration, post-dilution hemofiltration,
pre-dilution hemofiltration, pre-post-dilution hemofiltration or sequential hemodialysis.
4. The kidney substitution treatment machine according to one of the claims 1 to 3, wherein
the stored curves are concentration, absorbance or UV-absorbance of any waste product
present on the dialyzing fluid of any kidney substitution treatment.
5. The kidney substitution treatment machine according to one of the claims 1 to 4, wherein
the stored curves are KUV, single pool Kt/V or equilibrated Kt/V of any waste product
present on the dialyzing fluid of any kidney substitution treatment.
6. The kidney substitution treatment machine according to one of the claims 1 to 5, wherein
the stored curves are the reduction ratio of any waste product present on the dialyzing
fluid of any kidney substitution treatment, the single pool reduction ratio of any
waste product present on the dialyzing fluid of any kidney substitution treatment,
or the equilibrated reduction ratio of any taste product present on the dialyzing
fluid of any kidney substitution treatment.
7. The kidney substitution treatment machine according to one of the previous claims,
wherein the means to measure continuously any kidney substitution treatment related
waste product is a UV-Sensor.
8. The kidney substitution treatment machine according to one of the previous claims,
wherein the storage media is a hard disk hosted in the kidney substitution treatment
machine, a memory card hosted in the kidney substitution treatment machine or a patient
card attached to the kidney substitution treatment machine.
9. The kidney substitution treatment machine according to one of the previous claims
according to one of the previous claims, wherein the stored curves are normalized
to a given blood flow and/or a given clearance and/or a given dialysate flow.
10. The kidney substitution treatment machine according to one of the previous claims,
wherein the difference between therapies is quantified by an algorithm comparing the
urea under the curve of the considered curves.
11. The kidney substitution treatment machine according to one of the previous claims,
wherein one or more stored curves are shown simultaneously,
12. The kidney substitution treatment machine according to claim 11 wherein one best curve,
one worst curve and one average curve are stored and/or displayed.
13. A method for normalizing, displaying and storing curves describing the efficiency
of a kidney substitution treatment during said kidney substitution treatment,
wherein the kidney substitution treatment is provided by a machine, which has an extracorporeal
blood system pumping the patient blood at a preset blood flow rate through the blood
chamber of a dialyzer, which is divided by a semi-permcable membrane into the blood
chamber and a dialyzing fluid chamber and
wherein the dialyzing fluid flows at a preset flow rate through the dialyzing fluid
system of the machine and collects the waste products from the patient after flowing
through the dialyzing fluid chamber of the dialyzer and
wherein a device able to measure continuously any kidney substitution treatment related
waste product to deliver together with the data provided by the kidney substitution
treatment machine a adequacy parameter wherein the device is coupled with the dialyzing
fluid system of the kidney substitution treatment machine and wherein curves describing
the adequacy of the kidney substitution treatment are normalized to make them comparable,
and/or stored in an adequate media and/or displayed on the user interface of the machine
delivering the kidney substitution treatment and
wherein there are available means to calculate parameters quantifying the adequacy
differences between treatments.