[0001] The present invention relates to an optical signal noise cancellation system suitable
for use with optical sensor systems.
[0002] Several types of optical sensors have been developed for gas sensing applications,
utilizing direct optional absorption techniques.
[0003] One approach to gas sensors is a broad-band IR source-based CO
2 sensor for space applications. In this design, a broad-band IR source illuminates
a gas sample cavity and then impinges on two thermopile detectors. Optical interference
narrow-band IR filters are placed in front of the two thermopile detectors. The spectral
band for one of the filters is positioned within the CO
2 absorption band at around 4.2µm for direct optical absorption. The spectral location
of the other filter is located outside the CO
2 absorption band for reference purposes. The sensitivity of this device is about 1mm
Hg over a 0 to 30 mmHg PPCO
2 operating range. Several known problems hamper performance of this device: the sensor
exhibits significant errors (up to 11%) over a 40°C temperature range; sensor calibration
depends on system operation environment, gas mixtures, and total pressure, making
it less reliable for long-term applications in which these parameters may change;
due to lack of self-calibration and error checking capabilities inherent in the design,
the sensor calibration span is short; the sensor must be re-calibrated frequently
with operator intervention; and sensor output shows significant errors due to humidity
condensation in the IR absorption path during use.
[0004] The calibration dependence on gas mixture and total pressure for this device is inherent
in the sensing scheme. Although narrow-band IR filters are used to provide measurement
and reference signals, they are generally much wider than a number of spectral absorption
lines. The line-shape for these absorption lines will be collision-broadened as the
total gas pressure increases or as different gas components are added in mixture.
The narrow-band pass filters may not be able to completely integrate these broadened
line shapes resulting in significant integrated absorption dependence on pressure
and gas mixture as well as reference band signal variations.
[0005] Device temperature dependencies may arise from several factors. One possible source
is the temperature dependence of the broad-band IR source spectral output, producing
an unknown temperature dependence in the ratio signal output. The thermopile detectors
sense small temperature variation generated by IR energy. They are equally sensitive
to small environmental temperature changes that may generate difference in both detectors.
Since this sensor uses a DC sensing mode to resolve gas absorption signals, it is
highly susceptible to any contamination in the optical path, since the sensor can
not distinguish contamination-induced signal reduction from that of real absorption
signals. These factors dictate its short operation time span and the frequent needs
for system re-calibration.
[0006] Another class of gas analysis instrument consists of spectrophotometers using a broad-band
source, holographic gratings, and linear CCD arrays. A typical example is a fiber-optic
compact spectral analyzer marketed by Ocean Optics with typical spectral resolution
range of 1 to 2 nm. Due to size limitation imposed by space application requirements,
these diffraction-based devices do not offer enough resolution and sensitivity for
traces of gas analysis. Due to the broad-band source and low spectral resolution,
these devices suffer similar problems to those of the IR CO
2 type sensors. In addition, the light source and instrument construction does not
lend itself to a highly compact package design for space applications.
[0007] Diode laser-based spectral absorption techniques have been extensively studied and
are well-documented as a reliable, high-precision, high-sensitivity, and high-specificity
approach to measure gas species. Narrow-linewidth, spectral-pure (single-mode) diode
lasers are wavelength-tuned to resolve an entire spectral absorption line and accurately
measure absorption strength, corresponding to certain gas species. The progress of
diode laser technology in the past 10 years, driven by the telecommunication, data
storage, and consumer electronics industries, has led to reliable, spectrally-pure,
tunable, room- temperature diode lasers operating in several optical spectral windows
from 0.63 to 1.8 mm, suitable for probing a wide range of gas species. The life-time
of these devices are approaching 10
4 to 10
5 hours, consistent with advanced space exploration applications. Semiconductor diode
laser-based direct absorption sensors are most applicable for space environments since
they are miniature in size, have low-power consumption, and contain no-moving parts,
plumbing, gas handling, or high voltage components. They are capable of operating
in high vacuum or microgavity environments. The small mechanical and thermal mass
associated with diode lasers allows rapid wavelength tuning by current and temperature
to scan gas spectral absorption lines, resulting in rapid sensor response speed which
can satisfy a variety of monitoring, control, and warning systems requirements.
[0008] Several approaches to diode laser-based direct spectral absorption sensors have been
designed. The frequency modulation spectroscopy techniques require modulating laser
diodes to extremely high frequency, up to 1 Ghz, which is, in general, not practical
for sensor applications from a system design perspective. The wavelength modulation
spectroscopy techniques, on the other hand, modulate laser diodes with frequencies
up to MHz on top of a current sweep ramp in order to overcome diode laser noises.
This technique relies on phase sensitive detection to measure absorption line shapes.
The reported minimum measurable absorbance is on the order 10
-7, suitable for gas species trace measurements in many applications. However, the combination
of phase sensitive detection, near MHz modulation, and complicated absorption signal
extraction processes based on fundamental modulation frequency signal components and
various harmonics make such techniques less likely to be adapted in space applications,
where sensors must operate under severe environmental conditions, and multiple compensation
and sensor stabilization techniques must be applied. It is apparent that it is extremely
difficult to implement various system self-calibration, error-checking and compensation
approaches due to the complicated system requirements. Without these required elements
in place, it not likely that this type of device can satisfy space application requirements.
Additionally, the complicated electronics and signal processing requirements dictate
the use of very powerful microprocessors and associated elements in the final system
designs, which in general lead to higher power consumption, and a less favorable device
physical envelope for space applications.
[0009] U.S. Patent No. 5,134,276 discloses a detection scheme based on a balanced ratiometric
detector (BRD). A laser diode output beam is divided into measurement and reference
beams. The photocurrents generated by both laser beams are electronically balanced
via feedback-loop-controlled shunting of the reference beam photo-current, utilizing
a super-matched transistor pair in the circuit. Using this detection scheme to cancel
laser diode common-mode noises, research work demonstrated that absorbance on the
level of 10
-7 can be measured, when a laser diode is modulated at low frequencies (10 to 100 Hz).
Such low-frequency modulation of a laser-diode output leads to a typical tuning range
of 1 to 2 cm
-1 that allows full resolution of an atmospherically broadened absorption lineshape.
[0010] Certain disadvantages arising from various environmental factors must be properly
addressed to fully utilize the potentials of this technology. Some disadvantages include
the BRD circuit's logarithmic ratio signal which has an apparent linear temperature-dependence.
This problem may be solved by temperature regulation or compensation. However, when
required elements are examined for space applications, temperature regulation demands
additional power that is highly limited. Furthermore, the sensor will be more susceptible
to thermal shock induced errors due to the thermal recovery time and hysterisis of
the regulated elements. Temperature compensation will complicate signal processing
requirements and system design as each transistor pair used in the circuit may have
different temperature characterisitics. If the transisor pair operating point shifted
away from the linear region due to possible wide range of current loading, the temperature
dependence may become nonlinear, leading to more complications in compensation schemes.
Therefore, it is highly desirable that an improved BRD detection scheme be realized
that takes advantage of the demonstrated high-sensitivity and simplicity of the current
BRD design, while eliminating the apparent temperature sensitivity problems.
[0011] Another disadvantage is that the sensor system's self-calibration approaches for
long-term applications have not been developed. Optical-based sensor response characteristics
may change after initial calibrations due to factors such as laser aging, characteristics
shift of optical beam splitting components, detector characteristics variations, etc.
These are not unique problems for diode laser absorption gas species sensors, but
common and sometimes detrimental problems for many optical based sensor systems. Practical
implementation of optical-based sensors have been largely hampered by those problems,
as demonstrated by their slow acceptance in applications. The successful commercial
and military optical sensors to date all incorporate certain techniques to perform
self-calibration functions for system stabilization. Therefore, in order for the BRD-based
diode laser sensor to be successfully implemented in severe environmental condition
application with long calibration span and high reliability requirements, novel self-calibration
approaches must be developed to realize the potentials of this unique technology.
Ideally, the self-calibration techniques to be developed are applicable to a variety
of different gas species, having different sensitivity levels.
[0012] One of the major problems for direct spectral absorption diode-laser based sensors,
or any other optical sensors having an open optical path, is the humidity condensation
induced optical signal degradation in space life-support applications. In addition,
dirt may also build up in the open optical path during long-term applications. With
an intelligent compensation approach, the possible signal amplitude degradation caused
by the contamination will be interpreted as absorption signals, resulting in significant
errors. Without a well developed error compensation approach, these direct absorption
sensors may have short operation span and limited application range, as will similar
other optical and IR sensors having an open optical path. Therefore, an intelligent
error compensation approach must be developed in order to extend the BRD-based sensing
approach.
[0013] From a system design point view, it is important that a BRD-based sensor design examine
more closely additional issues such as overall system temperature sensitivity arising
not only from sensing schemes, but also from various sensor system components such
as diode lasers and detectors. For example, it is not practical to control laser diode
operating temperatures with very high precision in wide operating temperature ranges.
One must understand how the limited temperature-induced spectral shift of a diode-laser
impacts accuracy of the sensor systems.
[0014] It is noted that the above four aspects to further improve the diode-laser BRD-based
sensing scheme are generic system aspects, applicable to a variety of gas species
sensors. The successful development of these novel system self-calibration, stabilization,
and error compensation approaches will lead to a family of gas sensors suitable for
space and other commercial applications.
[0015] According to the present invention, there is provided an electronic signal noise
cancellation system suitable for use with optical sensor systems comprising:
a source of measurement current representing information signals impressed on a carrier
current signal that is modulated by undesirable noise signals;
a source of reference current representing said carrier current signal exclusive of
said information signals and modulated by said noise signals; and
combining means for additively combining said measurement current and a correction
current to substantially cancel the undesirable noise signals from said measurement
current and further produce a feedback control voltage which comprises an output signal
containing the information component of said measurement current;
characterised in that the system comprises a first voltage source generated proportional
to said reference current and a second voltage source (V
out) generated proportional to the first voltage source (V
ref) with a gain factor controlled by the feedback voltage, and the correction current
is generated by said second voltage source and modulated by the noise signals.
[0016] By examining the information signals individually and comparing the signals with
the original signals established at time of initial system calibration, various correction
factors can be calculated to perform system self calibration.
[0017] In the accompanying drawings:
FIGURE 1 is a block diagram of laser spectroscopy apparatus which is one application
of the present invention;
FIGURE 2 is a schematic diagram of exemplary circuitry that defines the invention;
FIGURES 3a-c are voltage waveforms associated with the invention;
FIGURE 4 is a method of providing dynamic scaling to correct for errors due to environmental
effects and is not an embodiment of the present invention.
FIGURE 5 is a graph of an absorption signal vs. total gas pressure and temperature.
[0018] The preferred embodiment of the invention will be illustrated as a noise suppression
system for an optical sensor for use in a gas sensor in a space environment. The gas
sensor application is for illustration purposes and is not intended to limit the use
of the invention.
[0019] The basic sensor system configuration is shown in Fig.1. A controller 19 sends control
signals to a Fabry-Perot (FB) or a Distributed Feedback (DFB) laser diode (0.76 µm
for O
2 sensing, 1.57 µm for CO
2 sensing ) 20 with proper temperature regulation. The output optical beam 22 is collimated
by collimating means 21 and is divided into a reference beam 26 and a measurement
beam 28 by a non-polarizing beam splitter 24. The measurement optical beam 23 passes
through a small multipass gas cell 30, providing sufficient path length for suitable
sensitivity to produce a modifed measurement beam 28'.
[0020] Referring also to Fig. 2, measurement beam 28' and reference beam 26 are input into
detector 38 and are converted into photocurrents I
mes and I
ref by photodiodes 32 and 34 respectively. Photocurrents I
mes and I
ref generate a ratio voltage signal V
c having laser common-mode excessive noise canceled, as described below. V
c is used to determine gas species density which is proportional to partial pressure
value.
[0021] An alternative sensor construction utilizes a 2x2 single-mode fiber-optic beam splitter.
The output of the laser diode is coupled into a small-core single-mode fiber ( <10
µm core diameter) by first collimating the beam to pass through an optical isolator,
and then focusing the beam into the single-mode fiber. The fiber construction offers
flexibility in alignment and in sensor system configurations. By using a nx2 fused
coupler, the fiber approach potentially allows multiplexing of several optical wavelengths
near 1.5 µm over one optical path in a sensor design,
[0022] Recognizing that both configurations using free standing optics or fiber coupled
optics may be advantageous for different purposes, the improved detector design and
proposed novel self-calibration techniques are applicable to both sensor configurations.
[0023] Referring to Fig. 2, I
mes is input via R6 and R10 to the inverting terminal of op amp 40 characterized as an
inverter gain stage amplifier. V
mes is the output of op amp 40 and is proportional to I
mes. Fig. 3a illustrates the waveform of V
mes which comprises two components. Component 100 is the linear element proportional
to the saw-tooth current ramp input to laser 20 discussed below. Component 110 represents
the signal due to the gas absorption.
[0024] I
ref is input to the inverting terminal of op amp 50 characterized as a transimpedance
amplifier. V
ref is the output of op amp 50 and is proportional to I
ref. Fig. 3b illustrates the waveform of V
ref.
[0025] V
in (equal to V
ref.) inputs into a feedback loop 56 via log-gain controlled amplifier 60 along with
control voltage V
c. Feed back loop 56 compises amplifier 60, op amp 62, resistor R4, op amp 70 characterized
as a trans impedance configuration, an integrator circuit comprising R7, C1 and op
amp 80. Amplifier 60 is controlled by the transfer function

where G and F are constants given by the temperature-compensated logarithmic amplifier
60 with a typical value of 0.5 and 0.25 respectively. As is obvious to those skilled
in the art, V
out is proportional to V
in with a gain factor which is controlled by the feedback loop where the gain factor
can be for example, linear, logarithic or any other function format. V
out is input into op amp 62 characterized as a buffer stage. V
out forces the voltage at node 2 to zero volts by changing the effective value of I
correction which is proportional to I
ref.
[0026] In operation, if I
mes > I
correction (voltage at node 2 ≠ 0), then a current input into op amps 70 and 80 causes V
c to increase and V
out to decrease resulting in I
correction to increase until voltage at node 2=0. The opposite relationship exists if I
mes < I
correction. Negative current flows from C1, causing V
c to decrease, through op amp 70 to node 1 causing and V
out to increase and reduce the value of I
correction until node 2 = zero volts.
[0027] Until feed back loop 56 forces I
mes = I
correction, current (either positive or negative) flows through amplifier 70 resulting in stored
voltage (either positive or negative) across C1 that is integrated to arrive at V
c. When photo-current balance is achieved at node 1 for canceling the laser common-mode
noises, the feedback loop control voltage V
c, as shown in Fig. 3c, relates to measurement and reference photocurrents I
mes and I
ref by the follow equation:

where the area 120 represents the signal relating to the volume of gas sensed in
chamber 30.
[0028] Feed back loop 56 may be tuned to provide a high response time in the µsec. range
or higher in response to changing values of I
mes or I
ref. The response time may be varied by adjusting R7 and C1 as is known to those skilled
in the art.
[0029] In an alternate embodiment, DC components not related to the measurement and reference
laser beam-generated signals 28' and 26 respectively, such as amplifier DC offset
and photodetector dark currents, can be effectively balanced using the bias resistor
pair R
8 and R
9, connected to a power supply V
s as a DC current sink. These DC components, if not properly removed, may lead to noticeable
signal baseline distortion and reduce the effectiveness of common mode noise cancellation.
During the calibration stage of detector 38, a bias voltage is added to node 2 by
adjusting the value of R8 to force current into or out from node 2 to cause the voltage
at node 2 to be zero, thereby, zeroing out the effects of the DC components relating
to component operating imperfections. As is readily apparent to those skilled in the
art, the bias voltage can be supplied by a microprocessor controlled digital-to-analog
converter that allows for continous calibration of the detector circuit 38.
[0030] With the embodiment of the invention shown in Fig. 2, the sensor as shown in Fig.
1 is capable of measuring absorption on weak vibrational overtone and combination
bands of targeted molecules. The absorption is described by Beer's Law:

where I
o,w is the initial laser intensity at frequency w, I
w is the transmitted intensity after propagation through a pathlength
I, S
T is the temperature dependent transition linestrength, N is the number density of
the absorbing molecules, and g
w is the Voigt line-shape function.
[0031] Referring again to Fig. 1, the diode laser 20 is thermally stabilized at an operating
temperature such that its output wavelength is close to a spectral absorption frequency
w of a gas component. The thermal tuning rate of diode lasers is on the order of 0.1
to 0.3 nm/°C, allowing access to multiple absorption lines. A periodic saw-tooth current
ramp is applied to the laser diode, where the self-heating of the p-n junction induced
by the input driving current produces small temperature changes that lead to similar
temperature-induced wavelength tuning. The low-frequency saw-tooth current ramp typically
produces a wavelength tuning of 1 to 2 cm
-1, allowing full resolution of an atomspherically broadened absorption line-shape.
V
mes and V
ref are depicted in Figs.3a-b, where the magnitude of the measurement signal attenuation
is determined by eq.(3). The logarithmic ratio output V
c shown in Fig.3c, contains the area 120 under the absorption Voigt lineshape which
is proportional to gas sample density.
[0032] V
c is input into a digital signal processor 27 where initial system calibration, a relation
or a numerical look-up table between the gas molecular density to that of the integrated
area under the log-ratio absorption signal in Fig.3c, is stored in the system memory.
The calibration table is established with initial system parameters such as laser
driving current ramp and temperatures, optical beam splitting ratio, detector and
circuit response values, etc. The sensor is then used to determine a gas components
density by comparing the measured integrated absorption signal to the calibration
table in the system memory. The partial pressure value can be directly calculated
using the measured gas component number density N and gas temperature T via the relation
P = NkT, where k is Boltzmann's constant.
[0033] After a calibrated sensor is placed in field use for an extended period of time various
parameters, such as temperature, humidy and dirt, may lead to the original system
calibration table to change, shifting system response characteristics away from that
given by the initial calibration table. Readout errors will result due to these unknown
changes.
[0034] Humidity effects fall into two categories: water vapor in the optical absorption
cavity and humidity condensation on optical components. Water vapor in the gas sample
cavity does not present problems to the spectral absorption sensing. For example,
for O
2 sensing, the strong absorption lines are at 760 nm, where nearby water vapor absorption
lines are around 820 nm. Likewise, for CO
2 sensing at around 1.57 µm wavelength, a group of spectral absorption lines having
absorption strength slightly greater than the O
2 lines at 760 nm can be readily separated from H
2O, CO absorption lines. Therefore, the presence of water vapor does not affect gas
component density measurement.
[0035] Humidity condensation presents a different set of problems, since it will scatter
and attenuate optical signals differently in the signal and reference paths. Without
compensation, such signal attenuation will be interpreted as signal absorption, leading
to significant errors. The problem is further complicated by the space operation environment,
such as spacesuit applications, where wide operating temperature range and high humidity
in human breathing air make the condensation more likely to occur. Similarly, since
the sensor has an open optical path, dirt may build-up on optical components for long-term
applications and cause similar signal attenuation. The dirt may result from, for example,
the volatile organic compounds in human breathing air after condensed moisture in
a sensor cavity is removed via evaporation.
[0036] A dynamical scaling method corrects for these error factors as shown in Fig. 4. An
assumption behind this method is that the gas molecule to be measured will maintain
its molecular form. The method consists of establishing a system reference for V
mes and V
ref based on initial system parameters including diode power output, optical beam splitting
ratio, optical path contamination, photodector response, amplifier response, etc.
91; storing an initial waveform for each V
mes and V
ref in system memory as calibration data tables 92; separating absorption line shape
120 from V
mes signal baseline signal and store both signal portions in system memory 93; periodically
monitor V
mes and V
ref and remove absorbtion line shape 120 from V
mes and compare signals to calibration data table 94; establishing a correction data
table for each V
mes and V
ref by comparing data in 94 with calibration data table generated in 92 above, 95; developing
a correction factor or correction table based on the ratio of correction data for
V
mes and V
ref 96; and appling correction factor or correction table to absorption line shape 120,
at step 97. The signals V
mes and V
ref represent a convoluted system opto-electronic-mechnical response characteristics.
The invention is capable of removing signal components not related to input optical
signals, as described above. Therefore, by comparing the non-absorption portion of
V
mes and V
ref with their initial calibration values stored in the system memory, self-calibration
factors in the form of scaling constants or a time-dependent function within a current
ramp for both signals can be established. The ratio of the correction factors is used
to correct for the logarithmic lineshape signal to calculate integrated absorption
areas. As such, the dynamical scaling-based self-calibration processes periodically
and actively measures the signal characteristics, and applies the correction factors
such that the system response will match the function established at the initial calibration
processes.
[0037] It is noted that this self-calibration approach is a generic process applicable to
any gas that can be measured with the diode laser spectral absorption technique.
[0038] The current ramp induced diode output wavelength scanning shown in Figs. 3a-c is
on the order of <0.1 nm. Therefore, humidity condensation and dirt build-up within
the optical path will appear to be spectrally flat within the absorption band of specific
gas component absorption lines. The signal attenuation resulting from such build-up
will proportionally reduce V
mes and the absorption lineshape while the reference signal will not be affected. Here,
only a correction factor for the measurement signal needs to be calculated using the
non-absorption portion of the baseline signal. This is actually a simplified case
of the self-calibration process. This process can be periodically applied or on demand.
Another benefit provided by the dynamical scaling process is that it will provide
a measure to determine system degradation such as laser diode output aging or extent
of dirt build-up in the optical path.
[0039] Test results indicate that common mode excessive noise can be reduced by 50 dB within
the operating bandwidth of the integrator feedback loop, which for the most part is
less than 100 kHz.
[0040] The dependence on temperature of the integrated absorption measurement area 120,
can be effectively eliminated by selecting an absorption wavelength with a very low
temperature-dependent line-strength. In the case of O
2, a number of lines within 760 to 761 nm and 763 to 764 nm, corresponding to the rotational
quantum numbers 11, 13, have an absorption line-strength temperature dependence on
the order of < 0.05% /°C. Fortunately, these lines are among the strongest absorption
lines of O
2, and are accessible with a number of commercially available FP or DFB laser diodes.
A group of CO
2 absorption lines around 1.57 µm exhibit similar behavior. These lines have slightly
stronger line strength than the 760 nm O
2 lines and they are clearly separated from other absorption lines for H
2O, CO, and other gas components. By measuring the integrated absorption strength at
these spectral locations, temperature-independent gas species number density N can
be determined. The partial pressure value can be calculated as p=NkT, where T is the
air temperature in the gas cell. In applications for which gas density alone is sufficient
to provide the necessary information for monitoring and control purposes, the sensor
system will be simplified since no air temperature sensor and associated electronics
are required.
[0041] The present diode laser direct absorption approach using detector 38, with the diode
modulated at low frequencies (10 to 100 Hz), is capable of resolving a pressure-broadened
line-shape. In the case of O
2 measurement, the individual spectral lines can be separated at a total pressure of
up to 75 psia, as verified by HITRAN data base. Experimental results show that the
spectral line-shape half-width increases almost linearly with pressure due to the
collision broadening processes; the integrated absorption area 120 stays the same.
One set of the data for O
2 measurement is shown in Fig. 5. In this experiment, 100 torr pure oxygen at a constant
temperature is released into a gas sample cavity. The integrated absorption signal
is measured and normalized to 1. Nitrogen gas was then slowly added to the gas cell
while maintaining the same temperature. As the total gas pressure increases at a constant
temperature, the integrated absorption signal remains constant with accuracies better
than 1%, as limited in most part by instrument uncertainties. This demonstrates that
the gas density measurement as well as the system calibration is independent of total
gas pressure and mixture. Therefore, this sensing approach provides reliable gas component
density and partial pressure measurement even as the total pressure and gas mixture
changes during applications. This also indicates that when the sensor system is calibrated
in one environment, it will be fully functional in other application environments
without the need for re-calibration. Additionally, the calibration processes can be
significantly simplified as only pure gas samples such as O
2 and CO
2 are required to perform the initial calibration, regardless of the gas composition
in applications. With such a simplified design consideration, a much simplified, portable,
field-service calibration module can be readily constructed to periodically calibrate
the sensor system. Coupled with the proposed self-calibration processes, it is expected
this system design will be able to maintain a high degree of long-term system stability
and data accuracy for a wide range of application conditions.
[0042] The integrated absorption signal also relates linearly to gas components in a linear
fashion as shown in our experiments ( better than 1% accuracy) for both O
2 and water vapor. This indicates that the sensor calibration look-up table can be
easily stored as a linear response function. In addition, during the necessary field
services, only one calibration point needs to be established, as we can use the system
self-calibration approach to re-build the calibration table while maintaining absolute
system accuracy levels.
[0043] The following table lists components and typical values which may be used to construct
the circuit shown in Fig. 2 for operation at a 100kHz bandwidth.
Table No. 1
| Component |
Designation |
Type or Value |
| Resistor |
R1, R4, R5 |
2,43K Ω |
| Resistor |
R2, R3, R10 |
4.99K Ω |
| Resistor |
R6, R7 |
10 Ω |
| Resistor |
R8 |
100K Ω |
| Resistor |
R9 |
50K Ω |
| Resistor |
R11 |
24.3K Ω |
| Capacitor |
C1 |
1500 pf |
| Op Amp |
40, 50, 62, 70, 80 |
AD817 |
| Log Gain Controlled Amplifier |
60 |
AD603 |
| Photo diodes |
32, 34 |
EG&G FFD-100 or FFD-200 |
[0044] It will be understood that the particular embodiments described above are only illustrative
of the principles of the present invention. The invention has application in any measurement
device which detects a small signal in combination with high noise content and further
utilizes a reference signal. Such applications may include acoustic, infared, frequency
or other types of measurement systems.